Recall of Device Recall SynchroMed II Implantable Infusion Pump

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Medtronic Neuromodulation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65443
  • Event Risk Class
    Class 1
  • Event Number
    Z-1579-2013
  • Event Initiated Date
    2013-06-03
  • Event Date Posted
    2013-06-25
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-11-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Pump, infusion, implanted, programmable - Product Code LKK
  • Reason
    Medtronic neuromodulation is providing healthcare providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they synchromed infusion pump. an electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und.
  • Action
    Medtronic sent an Urgent Medical Device Correction letter dated May 2013, to all affected customers. The letter described the issue of the pump's potentially shorting, how to identify if a pump is affected, and provided ongoing patient management recommendations. The Medtronic Rep would complete a Physician Notification Confirmation Form. A copy of the recall notification letter and the completed Confirmation Form is to be left with the physician. Customers with questions were instructed to call 1-800-707-0933. For questions regarding this recall call 763-526-9738.

Device

  • Model / Serial
    all SynchroMed II pumps
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    worldwide
  • Product Description
    Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II infusion system include the pump, catheter,r and catheter accessories || The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of drugs or fluids:
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Medtronic Neuromodulation, 7000 Central Ave NE, Minneapolis MN 55432-3568
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    USFDA