Recall of Plum 360 Infusion pump

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Hospira Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20993
  • Event Initiated Date
    2017-01-09
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
    Recalling Organisation: Hospira NZ Ltd, 58 Richard Pearse Drive, Airport Oaks, Mangere, Auckland 2022
  • Reason
    The manufacturer identified two software defects with the plum 360 infusion pump, software version 15.02 resulting in the following issues:, 1. the pump software incorrectly generates a depleted battery alarm instead of a replace battery alarm when the battery has a near end of life charge capacity, 2. when ac power is lost and quickly resumed, the pump software incorrectly classifies the event as an e323 alarm condition.
  • Action
    Instructions for use to be updated

Device

  • Model / Serial
    Model: , Affected: Software Version 15.02