Recall of GE Lightspeed CT Scanners with MDAS 16 5V power supply

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by GE Medical Systems LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18950
  • Event Initiated Date
    2015-08-03
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • Reason
    Ge healthcare has become aware that the service procedure used to replace a power supply on the rotating side of the gantry specifies the wrong torque for the mounting hardware. because of this error, the three mounting bolts securing the power supply could have been over-torqued when the component was replaced since installation. in the unlikely event that the component becomes loose on the ct gantry during operation, it could result in serious bodily injury if the component were to be expelled.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: LightSpeed Ultra, LightSpeed Plus, LightSpeed QX/i, LightSpeed 16, Affected: Manufactured from 2003-2006