Recall of Bard PowerPICC SOLO2 Catheterization kit, safety scapel

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Bard Access Systems.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17474
  • Event Initiated Date
    2014-10-06
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics, 54 Carbine Road, Mt Wellington, AUCKLAND
  • Reason
    There is a potential for an inadvertent scalpel stick injury involving the new safety scalpel which is included in the kit for users as for those accustomed to the previous scalpel two important differences are critical.
  • Action
    Instructions for use to be updated

Device

  • Model / Serial
    Model: Models: 1194118, 1295118, 1396118, 3194118, 3295118, 3396118, Affected: Lots shipped since 26-Mar-2014
  • Manufacturer

Manufacturer