Recall of MENISCAL KNOT PUSHER AND SUTURE CUTTER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SMITH & NEPHEW INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    40874
  • Event Risk Class
    II
  • Event Initiated Date
    2015-05-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A smith & nephew internal investigation found the packaging material and design to be insufficient in preventing the ultra fast fix knot pusher/cutter's sharp edges from puncturing the packaging pouch and rendering the device as non-sterile.

Device

  • Model / Serial
    Model Catalog: 72201537 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    MENISCAL KNOT PUSHER AND SUTURE CUTTER
  • Manufacturer

Manufacturer