Recall of ALARIS PATIENT CARE SYSTEM - PUMP

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CAREFUSION 303 INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    23566
  • Event Risk Class
    I
  • Event Initiated Date
    2010-09-07
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Pc unit model 8000 when used with pump or syringe modules may under certain conditions display an incorrect total dose when programmed to deliver bolus dose. the pump or syringe module delivers the actual programmed bolus dose (and associated volume to be infused) regardless of the displayed total bolus dose. this can cause confusion leading to improper treatment potentially causing injury or death.

Device

  • Model / Serial
    Model Catalog: 8000APM (Lot serial: >10 serial numbers.); Model Catalog: 8000APM (Lot serial: Contact manufacturer.)
  • Product Description
    Alaris PC Unit Model 8000
  • Manufacturer

Manufacturer