Recall of Medtronic MyCareLink Remote Monitors

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Medtronic Australasia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01009-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-08-10
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Medtronic is updating the instructions for use (ifu) for medtronic mycarelink remote monitors related to patients monitored on the medtronic carelink network with two (2) or more implanted medtronic heart devices. if a patient has multiple devices, there are potential impacts on the ability to remotely monitor that patient’s heart devices. these potential impacts could lead to missed carealert notifications or device reports. medtronic does not recommend the use of remote monitoring for patients with multiple implanted medtronic heart devices and clinicians should take this into account when implanting an additional medtronic heart device or preparing a patient for remote monitoring. for example, if a medtronic pacemaker is implanted in a patient with an existing reveal linqtm insertable cardiac monitor (icm), it may not be possible to monitor either device remotely.
  • Action
    Medtronic does not recommend the use of remote monitoring for patients with multiple implanted Medtronic heart devices and clinicians should take this into account when implanting an additional Medtronic heart device1 or preparing a patient for remote monitoring.

Device

  • Model / Serial
    Medtronic MyCareLink Remote Monitors Models: 24950, 25000ARTG Numbers: 215442, 283843 and 235570
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA