Guidant Corporation

12 devices in the database

  • Model / Serial
    Issue affecting devices is a pre-implant issue. Serial number ranges listed below may also include devices not affected (i.e. implanted devices).  Model H210 serial number 200031 thru 202462.  Model H215 serial numbers 700016 thru 702646.  Model H217 serial numbers 700006 thru 702543.  Model H219 serial numbers 200021 thru 201747.  Model H230 serial numbers 100012 thru 200526.  Model H235 serial numbers 100018 thru 650559.  Model H239 serial numbers 100018 thru 650272.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide, Puerto Rico, Virgin Islands and to countries that include: Austria, Belgium, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Iran, Ireland, Israel, Italy, Luxembourg, Mexico, Netherlands, Norway, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom
  • Product Description
    CONTAK RENEWAL 3 RF (models H210, H215), CONTAK RENEWAL 3 RF HE (models H217, H219), CONTAK RENEWAL 4 RF (models H230, H235) CONTAK RENEWAL 4 RF HE (model H239) cardiac resynchronization therapy defibrillator (CRT-D), provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricualar tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with suden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. The device also uses accelerometer-based adaptive-rate bradycardia therapy. These CONTAK RENEWAL RF devices feature enhanced telemetry communication with wandless ZIP telemetry. Cardioversion/defibrillation therapies include a range of low- and high-energy shocks using either a biphasic or monophasic waveform. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.
  • Model / Serial
    Model H170, serial numbers: 360449, 360632, 360655, 360657, 360661, 360666, 360714, 360719, 360721, 360738,  360741, 360742, 360743, 360745, 360746, 360749, 360756,  360758, 360763, 360766, 360790, 360791, 360792, 360796,  360800, 360802, 360844, 360847, 360848, 360861, 360863,  360904, 360907, 360908, 360910, 360986, 360987, 360988,  360989, 360991.  Model H175 serial numbers: 310338, 310434, 310523, 310532, 310533, 310537, 310544, 310545, 310706, 310707,  310709, 310713, 310719, 310721, 310734, 310735, 310737,  310746, 310753, 310796, 310798, 310813, 310907, 310908,  310909, 310913, 310916, 310976, 310990, 310996, 311007,  311010, 311011, 311013, 311015, 311016, 311092, 311096,  311152, 311154, 311181, 311185, 311386, 311389.   Model H177, serial numbers: 280992, 281011, 281012, 281017, 281020, 281021, 281022, 281024, 281045, 281056,  281066, 28107, 281077, 281119, 281121, 281123, 281124,  281126, 281127, 281143, 281175, 281213, 281233, 281247,  281320, 281356, 281364, 281368, 281371, 281376, 281385,  281401, 281403, 281413, 281476, 281480, 281484, 281485,  281491, 281533, 281539, 281542, 281544, 281559, 281562,  281580.   Model H179, serial numbers: 110905, 110907, 110908, 110950, 110952, 110954, 110955, 110958, 110980, 110983,  110987, 110988, 111000, 111001, 111003, 111004, 111007,  111012, 111024, 111041, 111042, 111128, 111173, 111180,  111198, 111201, 111248, 111250, 111254, 111263, 111294,  111299, 111308, 111313, 111334, 111335, 111336, 111338,  111340, 111341, 111342, 111346, 111350, 111352, 111390,  111391, 111392, 111393, 111397, 111422.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.
  • Product Description
    CONTAK RENEWAL 3 CRT-D (model no. H170, H175, H177, H179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 3 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.
  • Model / Serial
    Model H190, serial numbers: 310258, 310260, 310293, 310312, 310317, 310319, 310320, 310321, 310322, 310343,  310353, 310355, 310381, 310383, 310386, 310387, 310388,  310389, 310390.   Model H195, serial numbers: 350236, 350239, 350245,  350247, 350248, 350250, 350252, 350255, 350261, 350264,  350276, 350279, 350280, 350282, 350284, 350319, 350330,  350331, 350332, 350333, 350334, 350336, 350342, 350344,  350350, 350351, 350353, 350354, 350355, 350360, 350362,  350363, 350364, 350365, 350403, 350407, 350442, 350451,  350480, 350484, 350503, 350570, 350649.  Model H199, serial numbers: 320330, 320331, 320361, 320366, 320370, 320415, 320417, 320445.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.
  • Product Description
    CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.
  • Model / Serial
    Model T125, serial numbers: 115233, 115234, 115237, 115239, 115240, 115241, 115335, 115370, 115399, 115400,  115479.    Model T165, serial numbers: 110803, 110806, 110830, 110833, 110835, 110836, 111081, 111270, 111271, 111273,  111274, 111299, 111329, 111332, 111333, 111337, 111339,  111442, 111493, 111611, 111687, 111842, 111844, 111886,  111979, 112029, 112174, 112175.   Model T175, serial number 109787.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.
  • Product Description
    VITALITY DS (model T125) and VITALITY 2 (models T165, 175) ICD (implantable cardiac defibrillator)
  • Model / Serial
    Affected devices may be found in the following serial number ranges. INSIGNIA Entra SSI (model 0484) serial no. 100879 - 115237; (model 0485) serial no. 100088 - 100104; NEXUS Entra SSI (model 1326) serial no. 100122 - 100175. INSIGNIA Entra DDD (model 0985) serial no. 100059 - 100089; (model 0986) serial no. 100600 - 104439; NEXUS Entra DDD (model 1426) serial no. 100032 - 100091. INSIGNIA Entra SR (model 1195) serial no. 100739 - 101623; (model 1198) serial no. 102498 - 117468; NEXUS Entra SR (model 1398) serial no. 100067 - 100120. INSIGNIA Entra DR (model 1294) serial no. 101132 - 107445 & 292190; (model 1295) serial no. 100503 - 102668; (model 1296) serial no. 101519 - 153064; NEXUS Entra DR (model 1494) serial no. 100028 & 100029.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide and Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Polynesia, Portugal, Qatar, Reunion, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela
  • Product Description
    Guidant INSIGNIA I Entra family of pacemakers includes the following: SSI (model numbers 0484 and 0485); DDD (models 0985, 0986); SR (models 1195, 1198); and DR (models1294, 1295, 1296). Intermedics NEXUS Entra family of pacemakers includes the following: SSI (model 1326); DDD (model 1426); SR (model 1398) and DR (model 1494). The INSIGNIA I Entra pacemakers are multiprogrammable pacemakers from Guidant. The NEXUS I Entra pacemakers are multiprogrammable pacemakers from Intermedics. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. The INSIGNIA and NEXUS I Entra adaptive-rate models have an accelerometer, which is a motion sensor that responds to patient activity. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. or Intermedics, Inc., 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.
7 more

13 manufacturers with a similar name

Learn more about the data here

  • Manufacturer Address
    Guidant Corporation ACS, 26531 Ynez Road, Temecula CA 92591-4628
  • Source
    USFDA
  • Manufacturer Address
    Guidant Corporation Advanced CardiovascularSystem, 26531 Ynez Rd, Temecula CA 92591-4630
  • Source
    USFDA
  • Manufacturer Address
    Guidant Corporation, 4100 Hamline Ave N, Saint Paul MN 55112-5700
  • Source
    USFDA
  • Manufacturer Address
    Abbott Vascular-Cardiac Therapies dba Guidant Corporation, 26531 Ynez Rd, Temecula CA 92591-4630
  • Source
    USFDA
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.” Note published by the authorities from the Netherlands: This message is a warning from the manufacturer. After placing a medical device on the market, the manufacturer is obliged to follow the device and, where necessary, to improve it. When there is a risk for the user, the manufacturer must take action. The manufacturer informs users and the inspection of this action with a warning (Field Safety Notice). See also medical device warnings. This is part of the supervision of medical technology.
  • Source
    IGJ
8 more