Recall of Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31232
  • Event Risk Class
    Class 2
  • Event Number
    Z-0864-05
  • Event Initiated Date
    2005-02-09
  • Event Date Posted
    2005-06-02
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-06-29
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Orthopedic Manual Surgical Instrument - Product Code LXH
  • Reason
    The engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
  • Action
    The firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.

Device

  • Model / Serial
    Chisel lot number 60048084.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Nationwide and Canada.
  • Product Description
    Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer Inc., 345 E Main St, Warsaw IN 46580-2746
  • Source
    USFDA