Recall of VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31719
  • Event Risk Class
    Class 3
  • Event Number
    Z-0913-05
  • Event Initiated Date
    2005-03-15
  • Event Date Posted
    2005-06-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-02-10
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Enzyme Immunoassay, Non-Radiolabeled, Total Thyroxine - Product Code KLI
  • Reason
    Potential for positively biased results in qc and patient samples.
  • Action
    Letters dated 3/15/2005 sent First Class Mail. Customers are advised product will be replaced. The letter includes modified quality control procedures that will pre-qualify each reagent pack.

Device

  • Model / Serial
    Lot 710, Exp. 7/28/05
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide distribution. Product shipped worldwide via J&J; foreign affiliates.
  • Product Description
    VITROS Immunodiagnostic Products TOTAL T4 REAGENT PACK, Ref #874 4468, 100 coated wells. Firm on label: Ortho-Clinical Diagnostics, Amersham UK.