Recall of Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27905
  • Event Risk Class
    Class 2
  • Event Number
    Z-0298-04
  • Event Initiated Date
    2003-11-26
  • Event Date Posted
    2004-01-06
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2004-02-20
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Fluorometric Method, Cpk Or Isoenzymes - Product Code JHX
  • Reason
    Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
  • Action
    Customers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.

Device

  • Model / Serial
    Lot 611, Exp. 25 June 2004
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.
  • Product Description
    Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ortho-Clinical Diagnostics, 100 Indigo Creek Drive, Rochester NY 14650-0881