Recall of Vitros Immunodiagnostic HBsAg CONFIRMATORY KIT * REF 680 1324. Firm on label: Ortho-Clinical Diagnostics, 62 The Broadway, Amersham, Bucks, UK. UPC *+H221680132412* . Each kit contains: 1 x kit of reagents, and protocol card for HBsAg Confirmatory Kit. According to the firm (but not on label), there are 60 neutralization reactions per confirmatory kit.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    34238
  • Event Risk Class
    Class 1
  • Event Number
    Z-0327-06
  • Event Initiated Date
    2005-12-15
  • Event Date Posted
    2006-01-06
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-08-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm) - Product Code LOM
  • Reason
    Increased background signal of the sample diluent in the affected confirmatory kits could cause some results to be classified as 'not confirmed'.
  • Action
    Letters sent overnight mail on 12/15/2005. Customers instructed to discontinue use and discard remaining inventory. In a separate Q&A; sheet, Ortho recommends that previously reported results be reviewed.

Device

  • Model / Serial
    Lot 5140, exp. 4/13/2006; Lot 5150, exp. 5/31/2006; and Lot 5160, exp. 7/3/2006
  • Product Classification
  • Device Class
    3
  • Implanted device?
    No
  • Distribution
    Nationwide distribution to end users (medical facilities, testing labs and public health agencies). One J&J; foreign affiliate in Puerto Rico with sub-distribution in the Caribbean.
  • Product Description
    Vitros Immunodiagnostic HBsAg CONFIRMATORY KIT * REF 680 1324. Firm on label: Ortho-Clinical Diagnostics, 62 The Broadway, Amersham, Bucks, UK. UPC *+H221680132412* . Each kit contains: 1 x kit of reagents, and protocol card for HBsAg Confirmatory Kit. According to the firm (but not on label), there are 60 neutralization reactions per confirmatory kit.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ortho-Clinical Diagnostics, 100 Indigo Creek Dr, Rochester NY 14626-5101
  • Manufacturer Parent Company (2017)
  • Source
    USFDA