Recall of Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30570
  • Event Risk Class
    Class 3
  • Event Number
    Z-0387-05
  • Event Initiated Date
    2004-11-22
  • Event Date Posted
    2005-01-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-05-02
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    C-Reactive Protein, Antigen, Antiserum, And Control - Product Code DCK
  • Reason
    Results may be negatively biased due to a non-linear response at the upper end of the reportable (dynamic) range.
  • Action
    Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.

Device

  • Model / Serial
    Coatings 401 thru 419.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
  • Product Description
    Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ortho-Clinical Diagnostics, 100 Indigo Creek Dr, Rochester NY 14626-5101
  • Manufacturer Parent Company (2017)
  • Source
    USFDA