Recall of Vitros Chemistry CRBM Slides, REF 889 2382. There are 90 slides per box. For in vitro diagnostic use. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32303
  • Event Risk Class
    Class 2
  • Event Number
    Z-1016-05
  • Event Initiated Date
    2005-05-26
  • Event Date Posted
    2005-07-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-06-12
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Enzyme Immunoassay, Carbamazepine - Product Code KLT
  • Reason
    Falsely elevated carbamazepine results.
  • Action
    Letters dated 5/26/05 with instructions to discontinue use and discard slides. Customers also instructed to review previous results with Laboratory Medical Director to determine need for further consultation with health care provider.

Device

  • Model / Serial
    Generations 18, 19, 20, 21, and 23
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    423 U.S. consignees nationwide, and 22 OCD/J&J; foreign affiliates (Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Peru, Ecuador, India, Japan, Mexico, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain)
  • Product Description
    Vitros Chemistry CRBM Slides, REF 889 2382. There are 90 slides per box. For in vitro diagnostic use. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626.