Recall of Varian

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Varian Medical Systems Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    45520
  • Event Risk Class
    Class 2
  • Event Number
    Z-0544-2008
  • Event Initiated Date
    2007-09-10
  • Event Date Posted
    2008-02-14
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-11-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Linear Accelerator - Product Code IYE
  • Reason
    Treatment error: the collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle.
  • Action
    On 9/10/07, the firm issued an Urgent Medical Device Correction letter to all its direct consignees, informing them of the device defect and providing instructions on the recall. Short term user instructions were provided and the firm will schedule follow-up visits to make the needed permanent device corrections at each device site.

Device

  • Model / Serial
    The Serial Numbers: H271306; including: H140010, H140046, H140047, H140060, H140095, H140098, H140120, H140124, H140130, H140143, H140144, H140145, H140148, H140152, H140201, H140202, H140203, H140204, H140208, H140209, H140210, H140211, H140213, H140214, H140217, H140218, H140220, H140221, H140223, H140224, H140225, H140226, H140227, H140229, H140233, H140235, H140237, H140703, H140706, H140715, H140716, H140721, H140723, H140725, H140727, H140728, H140731, H140735, H140742,  H270144, H270357, H270412, H270423, H270518, H270608, H270615, H270618, H270680, H270707, H270750, H270792, H270811, H270845, H270862, H270885, H270890, H270893, H271013, H271023, H271031, H271049, H271112, H271113, H271118, H271122, H271154, H271169, H271224, H271275, H271333, H271341, H271345, H271358, H271365, H271413, H271444, H271602, H271673, H271790, H271852, H271879, H271890, H271893, H271906, H271940, H271962, H271964, H271977, H271999, H272023, H272061, H272105, H272108, H272148, H272153, H272184, H272203, H272230, H272241, H272242, H272273, H272319, H272320, H272360, H272375, H272380, H272437, H272452, H272494, H272804, H272822, H272854, H272856, H272892, H272929, H272931, H272982, H273001, H273007, H273019, H273042, H273086, H273120, H273123, H273127, H273132, H273136, H273137, H273146, H273157, H273165, H273180, H273189, H273190, H273197, H273209,  H290106, H290107, H290142, H290165, H290176, H290179 , H290186, H290201, H290203, H290205, H290206, H290213, H290215, H290236, H290244, H290254, H290259, H290326, H290358, H290404, H290503, H290504, H290508, H290512, H290517, H290519, H290520, H290535, H290548, H291002, H291005, H291008, H291012, H291014, H291017, H291018, H291023, H291024, H291025, H291033, H291036, H291037, H291043, H291046, H291047, H291050, H291053, H291056, H291059, H291062, H291064, H291069, H291070, H291071, H291073, H291082, H291083, H291084, H291086, H291087, H291091, H291094, H291095, H291096, H291099, H291108, H291109, H291112, H291113, H291117, H291119, H291121, H291127, H291128, H291129, H291130, H291133, H291135, H291140, H291141, H291144, H291150, H291158, H291163, H291164, H291165, H291166, H291167, H291168, H291169, H291170, H291171, H291172, H291174, H291178, H291179, H291181, H291182, H291185, H291193, H291196, H291200, H291203, H291204, H291205, H291206, H291208, H291209, H291211, H291214, H291216, H291217, H291218, H291220, H291223, H291224, H291231, H291237, H291238, H291241, H291247, H291248, H291250, H291251, H291254, H291255, H291256, H291264, H291265, H291269, H291270, H291272, H293303, H293311, H293313, H293319, H293330, H293331, H293335, H293342, H293344, H293346, H293348, H293355, H293356, H293360, H293367, H293385, H293394, H293395, H293397, H293398, H293400, H293401, H293404, H293405, H293406, H293410, H293414, H293416, H293417, H293418, H293423, H293425, H293426, H293437, H293438, H293439, H293441, H293447, H293449, H293451, H293452, H293456, H293457, H293458, H293466, H293470, H293475, H293476, H293477, H293478, H293479, H293480, H293481, H293483, H293484, H293488, H293490, H293502, H293505, H293509, H293515, H293516, H293517, H293518, H293525, H293531, H293537, H293541, H293543, H293544, H293546, H293557, H293558, H293560, H293561, H293562, H293565, H293572, H293581, H293583, H293593, H293596, H293598, H293599, H293614, H293627, H293630, H293635, and H293642
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide- Canada, Mexico, and the USA
  • Product Description
    Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system: Model: High Energy Clinacs with one or more of the following options: || -- BrainLAB micro MLC, || -- 6MV SRS photon beam, || -- Fine Beam Isocenter Accuracy, || -- Stereotactic motion disable. || ----Product is manufactured and distributed by Varian Medical Systems, Palo Alto, CA
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Varian Medical Systems Inc, 911 Hansen Way, Palo Alto CA 94304
  • Source
    USFDA