Recall of UniCel DxI and DXC Immunoassay Systems

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62706
  • Event Risk Class
    Class 2
  • Event Number
    Z-2259-2012
  • Event Initiated Date
    2011-10-13
  • Event Date Posted
    2012-08-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-08-23
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Reason
    The recall was initiated because beckman coulter has received seven reports of a software anomaly for the unicel dxl immunoassay systems. if undetected, this anomaly may cause erroneous test results.
  • Action
    The firm, Beckman Coulter, sent an "URGENT PRODUCT CORRECTION" letter dated October 13, 2011 with attached PCA Response Form to all customers who purchased the UniCel Dxl Immunoassay Systems and DxC Clinical Systems which includes all software versions. The letter describes the product, problem and actions to be taken. The customers were instructed verify the system temperatures immediately following each initialization; do not begin processing samples until this verification is complete; refer to the appendix included with the letter for detailed instructions about verifying system temperatures; review their quality control (QC) performance history; and share this information with their staff and retain this information as part of their quality system documentation, and to complete and return the enclosed response form within 10 days to Beckman Coulter, Inc., 1000 Lake Hazeltine Drive, Chaska, MN 55318; Attn: Regulatory Affairs or fax to: 1-786-639-4000. Customers with any questions regarding the notice, were instructed to contact Customer Support Center via our website, http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter Representatives.

Device

  • Model / Serial
    Part Numbers:  DxC 660i: A64871;  DxC 680i: A64903; DxC 860i: A64935; DxC 880i: A59102
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide distribution: USA (nationwide) and countries of: Algeria, Angola, Argentina, Australia, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea,Republic of; Kuwait, Liechtenstein, Malaysia, Mayotte, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Philippines, Poland, Puerto Rico, Romania, Russian Federation, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.
  • Product Description
    UniCel DxC 660i, 680i, 860i, 880i SYNCHRON Access Clinical Systems, || Part Numbers: || DxC 660i: A64871; || DxC 680i: A64903; || DxC 860i: A64935; || DxC 880i: A59102. || The DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Beckman Coulter Inc., 250 S Kraemer Blvd, Brea CA 92821-6232
  • Manufacturer Parent Company (2017)
  • Source
    USFDA