Recall of The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Arrow International Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31765
  • Event Risk Class
    Class 2
  • Event Number
    Z-0842-05
  • Event Initiated Date
    2005-03-18
  • Event Date Posted
    2005-05-13
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-11-22
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Catheter, Embolectomy - Product Code DXE
  • Reason
    The orange pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
  • Action
    The firm began notification of consignees on 3/23/2005 via certified letter.

Device

  • Model / Serial
    All units with Lot Numbers beginning with: MF311; MF312; MF401; MF402; MF403; MF404; MF405; MF406; MF407; MF408; MF409; MF410; MF411; MF412; MF501.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Product was distributed nationwide to dealers, hospitals, sales representatives, and subsidiaries (distributors). The product was also distributed to Australia, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Japan, Netherlands, Russia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey, United Kingdom. There are no government accounts.
  • Product Description
    The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Arrow International Inc, 2400 Bernville Rd, Reading PA 19605-9607
  • Manufacturer Parent Company (2017)
  • Source
    USFDA