Recall of Terumo APS 1

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Terumo Cardiovascular Systems Corp.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    37244
  • Event Risk Class
    Class 3
  • Event Number
    Z-0924-2007
  • Event Initiated Date
    2004-04-01
  • Event Date Posted
    2007-06-30
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-06-11
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Perfusion System - Product Code DTQ
  • Reason
    A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
  • Action
    The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.

Device

  • Model / Serial
    Serial numbers 0033 through 0355 are potentially affected by the problem. Serial numbers 0033, 0041, 0049, 0052, 0105, 0115, 0117, 0119, 0122, 0145, 0151, 0159, 0160, 0161, 0162, 0170, 0174, 0218, 0226, 0242, 0257, 0262, 0277, 0279, 0290, 0299, 0300, 0305, 0306, 0331, 0334, 0339, 0343 and 0360 have been corrected.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
  • Product Description
    Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Terumo Cardiovascular Systems Corp, 6200 Jackson Rd, Ann Arbor MI 48103-9586
  • Source
    USFDA