Recall of Roche brand ISE SnapPak for use only in the AVL 9180/9181 Electrolyte Analyzer Assembled and distributed by Roche Diagnostics Corp., Indpls., IN.; Ref/No. 03112349180, Part number BP5186.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30664
  • Event Risk Class
    Class 2
  • Event Number
    Z-0606-05
  • Event Initiated Date
    2004-12-21
  • Event Date Posted
    2005-03-11
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-06-30
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel data.
  • Extra notes in the data
  • Reason
    A defective seam in the two chamber pouch may result in the mixing of reagents, which would either result in an inability to calibrate the analyzer or would bias patient results.
  • Action
    Own name distributor consignees were notified on or about 1/3/05 via letter dated 12/21/04, and were requested to discontinue shipment of the product, notify all of their customers of the recall and to discard product. Roche consignees were notified via letter dated 12/21/04, requested to notify their customers if they had further distributed the product, and instructed to discard all inventory of product.

Device

  • Model / Serial
    All units.
  • Product Classification
  • Distribution
    Nationwide, Argentina, Austria, Brazil, Chile, China, Colombia, Ecuador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Panama, Philippines, Poland, Russia, San Salvador, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Uruguay.
  • Product Description
    Roche brand ISE SnapPak for use only in the AVL 9180/9181 Electrolyte Analyzer Assembled and distributed by Roche Diagnostics Corp., Indpls., IN.; Ref/No. 03112349180, Part number BP5186.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Roche Diagnostics Corp., 9115 Hague Rd, Indianapolis IN 46256-1025
  • Source
    USFDA