Recall of Regenerex Ringloc Solid Acetabular Cup

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Biomet, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    58663
  • Event Risk Class
    Class 2
  • Event Number
    Z-2449-2011
  • Event Initiated Date
    2011-04-14
  • Event Date Posted
    2011-06-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-09-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, hip, semi-constrained (metal, uncemented acetabular component) - Product Code LPH
  • Reason
    The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. if the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
  • Action
    Biomet Orthopedics sent an URGENT MEDICAL DEVICE RECALL NOTICE DATED April 14, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Immediately locate and remove the affected product. Follow the instructions on the enclosed "FAX Back Response Form" and fax a copy of the Response Form to 574-372-1683 prior to return of product. Customers were to confirm receipt of the notice by calling 800-348-9500, ext 3755 or 3756. For any questions regarding this recall call 574-371-3755 or 574-371-3756.

Device

  • Model / Serial
    Lot number M817580.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution - USA (nationwide) and the countries of Australia, Netherlands and Japan.
  • Product Description
    Regenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 52 mm Size 23, w/Lock Ring w/Plug, sterile, REF PT-104052, Biomet Orthopedics, Warsaw, IN. || Intended Use: The Regenerex Ringloc is intended to replace the natural acetabulum damaged by disease, trauma or revision of previous arthroplasty. Indications For Use: 1 . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis 3. Correction of functional deformity 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Biomet, Inc., 56 E Bell Dr, Warsaw IN 46582
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA