Recall of reference electrodes

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Radiometer America Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    35788
  • Event Risk Class
    Class 3
  • Event Number
    Z-1323-06
  • Event Initiated Date
    2005-09-19
  • Event Date Posted
    2006-08-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-02-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    reference electrodes - Product Code CHL
  • Reason
    Reference electrodes for the firm's abl 700/800 blood gas analyzers are oversized and will not fit into the instrument.
  • Action
    Following the receipt of three (3) customer complaints, on 9/14/2005, the firm issued an E-Mail notification [Management Cover] to their Service Personnel on 9/19/2005. The Management Cover instructed the service personnel to check the electrodes located at their stock and their customer locations; put all electrodes, which cannot be inserted in their call-back stock; and provide for credit to their customers through normal channels. The Service Personnel in turn, contacted their customers via telephone on or around 9/20/2005 and asked that they check their inventories for the presence of any electrodes which would not fit in the analyzer and return any such electrodes to Radiometer America for replacement

Device

  • Model / Serial
    Part #945-603 - Lot #96-01 and #96-02
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide and Canada.
  • Product Description
    Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number: 945-603, distributed one (1) electrode per box.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Radiometer America Inc, 810 Sharon Dr, Westlake OH 44145-1598
  • Source
    USFDA