Recall of PowerLED 700 Surgical Light System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Maquet Medical Systems USA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    68142
  • Event Risk Class
    Class 2
  • Event Number
    Z-1730-2014
  • Event Initiated Date
    2014-04-16
  • Event Date Posted
    2014-06-06
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-04-06
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Light, surgical, ceiling mounted - Product Code FSY
  • Reason
    Maquet has identified an issue with the maquet powerled 700 (certain serial numbers). one or several plastic fixation tabs can break, which can cause the ambient light module to partially detach.
  • Action
    Maquet issued "Urgent-Medical Device Field Correction" notices, dated April 16, 2014 to affected customers via Federal Express. The notice identified the product, related issue and the action needed to be taken by the customer. The corrective action will include inspection of the fixation tabs of the Ambient Light Module adn replacement of the Ambient Light Module if tabs are found to be defective. The correction will be performed by an authorized MAQUET Service Technician. MAQUET Service Team representatives will schedule on-site service with customers. If customers have any questions or require additional information, the notice instructs them to contact their local MAQUET representative, or MAQUET Customer Service at 1-888-627-8383 (option 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.

Device

  • Model / Serial
    Part number (begins with ALM) and associated serial numbers:  ALM568300999 13, 14, 16, 19, 20, 21, 22, 23, 24, 25, 29, 30, 31, 32, 34, 35, 36, 41, 50  ALM568301999 13, 14, 16, 17, 20, 23, 24, 25, 26, 52, 54, 55, 56, 57, 60, 63, 64, 65, 66, 70, 71, 72, 74  ALM568302999 11, 43, 44, 46, 49, 50, 51, 54, 57, 58, 59, 6, 61, 63, 64, 66, 67, 71, 72, 74, 77, 78, 80, 81  ALM568303999 8, 9, 32, 52, 53, 58, 60, 61, 62, 65, 66, 69, 70, 71, 73, 74, 75, 77, 78, 79, 80, 81, 82, 85, 86, 87, 89, 91, 92, 94, 95, 96, 97, 99, 100, 105, 106, 108, 111, 118, 120, 121, 122, 159  ALM568320901 12012, 40015, 40017, 40018, 40019, 40020, 40021, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 50022, 50023, 50024, 50025, 50026, 50027, 50028, 50030, 50031, 50032, 50033, 50034, 50035, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50047, 50048, 50050, 50051, 50052, 50053, 50054, 50055, 50056, 50058, 50059, 50060, 50061, 50064, 50066, 50067, 50068, 50069, 50070, 50071, 50072, 70033, 70034, 70035   ALM568320905 40003, 40004, 40005, 40006, 40007, 40008, 40009, 40010, 40011, 40012, 40013, 40014, 40015, 40016, 50003, 50004, 50005, 50006, 50008, 50009, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 60003, 60004, 70001, 70002, 70003, 70004, 70005, 70006, 70007, 70009, 70010, 70011, 70012, 70013, 70014, 70015, 70016, 70017, 70018, 70019, 70020, 70021, 70022, 70024, 70025, 70027, 70028, 70029, 70030, 70031, 70032, 70033, 70034, 70035, 70036, 70037, 70038, 70039, 70040, 70041, 70042, 70043, 70044, 70045, 70046, 70047, 70048, 70049, 70050, 70051, 70052, 70053, 70054, 70055, 70056, 70057, 70058, 70059, 70060, 70061, 70063, 70064, 70066, 70068, 70069  ALM568370900 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20039, 20040, 20041, 20042, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20054, 30038, 30039, 30053, 30054, 40006, 40010, 40012, 40013, 40014, 40015, 40016, 40018, 40019, 40020, 40027, 40028, 40029, 40030, 40031, 40032, 40033, 40034, 50000, 50001, 50004, 50006, 60007, 60008, 60009, 60010, 60012, 60031, 60041, 60042, 60043, 60044, 60046, 60047, 60048   ALM568370901 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20038, 20039, 20040, 20041, 20042, 20045, 20053, 20054, 20055, 20056, 20057, 20058, 20059, 20060, 20061, 20062, 20064, 20065, 20066, 20067, 20068, 20069, 20070, 32042, 40028, 40033, 40034, 40035, 40036, 40037, 40038, 40045, 50002, 50006, 50008, 50009, 50024, 50035, 50036, 50037, 50039, 50040, 50041, 50042, 50043, 50044, 50045, 50046, 50047, 50048, 60030   ALM568370902 20011, 20015, 20016, 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20040, 20041, 20042, 20048, 20049, 20050, 20051, 20057, 20058, 20059, 20060, 20061, 20062, 20063, 20064, 20067, 20068, 20069, 20070, 20071, 20072, 20073, 20074, 20075, 20076, 20077, 20078, 20079, 20080, 20083, 20086, 20087, 20089, 20090, 20091, 20094, 20095, 20096, 20098, 20104, 20105, 20106, 20107, 20108, 20109, 20110, 20111, 20112, 20113, 20114, 20115, 20116, 20117, 20118, 20119, 20120, 20121, 20122, 20123, 20124, 20125, 20127, 20132, 20133, 20135, 20136, 20137, 20139, 20140, 20141, 20144, 20151, 20152, 30000, 30051, 30053, 30054, 30062, 30063, 30065, 30066, 30067, 30068, 30078, 30081, 30090, 30091, 30092, 30099, 30107, 30127, 30128, 30130, 30131, 30134, 32164, 40000, 40006, 40007, 40009, 40011, 40012, 40013, 40014, 40015, 40017, 40018, 40019, 40020, 40021, 40023, 40039, 40040, 40041, 40042, 40043, 40044, 40045, 40046, 40047, 40048, 40050, 40051, 40052, 40053, 40054, 40055, 40056, 40057, 40058, 40060, 40064, 40065, 40067, 40078, 40081, 40082, 40083, 40084, 60030, 60031, 60036   ALM568370903 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20041, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20055, 20056, 20057, 20058, 20059, 20066, 20067, 20072, 20073, 20074, 20075, 20076, 20077, 20079, 20080, 20081, 20082, 20084, 20087, 20088, 20089, 20090, 20091, 20092, 20093, 20097, 20098, 20099, 30072, 40025, 50000, 50001, 50002, 50003, 50004, 50006, 50028, 50051, 50052, 50054, 50055, 50057, 50058, 50059, 50060, 60024, 60025, 60026, 60028  ALM568370904 20011, 20013, 20014, 20015, 20016, 20017, 20018, 20019, 20020, 20023, 20024, 20025, 20026, 20027, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 30000, 30001, 30002, 30003, 30004, 30012, 30013, 30014, 30015, 30018, 30019, 30020, 30025, 30026, 30027, 30028, 30029, 30030, 30031, 30032, 30033, 30034, 30035, 30036, 30037, 30038, 30039, 30040, 30041, 30042, 30043, 40011, 40012, 40013, 40014, 40015, 40016, 40017, 40022, 40023, 40024, 40031, 40032, 40033, 40034, 40035, 40036, 50004, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50019  ALM568370905 20010, 20016, 20017, 20018, 20019, 20022, 20023, 30027, 40012, 40022, 40023, 40024, 40025, 40027, 40028, 40029, 40031, 40032, 40033, 40037, 40040, 40041, 40047, 40049, 40050, 50005, 50006, 50012, 50015, 50018, 50021, 60007, 60023, 60027, 60030, 60032, 60033
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US Nationwide Distribution.
  • Product Description
    Maquet PowerLED 700. || Product Usage; To provide illumination to the surgical area or the patient.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Maquet Medical Systems USA, 45 Barbour Pond Dr, Wayne NJ 07470-2094
  • Manufacturer Parent Company (2017)
  • Source
    USFDA