Recall of PercuNav Field Generator Stand

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips Medical Systems North America Co. Phillips.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    58429
  • Event Risk Class
    Class 2
  • Event Number
    Z-2219-2011
  • Event Initiated Date
    2011-04-01
  • Event Date Posted
    2011-05-19
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-03-15
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Computed Tomography X-Ray System - Product Code JAK
  • Reason
    The percunav field generator stand could tip over during set up, during use, or during transport.
  • Action
    On 4/4/11, Philips Healthcare began sending out the URGENT-Medical Device Correction notification letter to their consignees. Philips informed their customers of a potential tipping hazard of the PercuNav Field Generator Stand. The notification letter also provides instructions to customers to ensure the articulating Mounting Arm is positioned so the Field Generator is not further than the specified distance from the back of the clamp on the stand's central pole during set up or during use. Customers are instructed to apply a provided label on stand to warn of a potential for tipping due to excessive force. For transport, customers are instructed to remove the Field Generator from the Mounting Arm and place the Field Generator in the storage compartment on the system cart. The Mounting Arm should then be collapsed and folded into a 90 degree angle. Customers are informed that a Philips field service representative will contact them to schedule installation of a service kit on the stand, free of charge, once service kits are available. Customers can call Philips Customer Service at 1-866-767-7822 and reference to FCO 79500197 for any questions about this recall.

Device

  • Model / Serial
    System Code Serial Numbers:  03B8Y3, 20071000005, 20071000006, 20071000007, 20071000008, 20071000009, 20071000010, 20081000011, 20081000013, 20091000019, 20091000020, 20091000021, 20091000022, 20091000023, 02RCHF, 02X3NK, 034K1R, 039VDK, 039W1H, 03B4F3, 03B59R, 03B8MW, 03B8N0, 03B8QR, 03B8XX, 03B8XY, 03B8XZ, 03B8Y0, 03B8Y1, 03B8Y4, 03B8Y5, B0091T, B009BJ, B009HV, B009VJ, B00DDJ, B00DQ2, B00DQX, B00DQZ, B00DRF, B00DRP, B00DT7, B00DXM, B00DY0, B00F1Q, B00F3H, B00F3M, B00HKW, B00HKZ, B00HV4, B00HV7, B00HW3, B00HZR, B00J05, B00P3X, B00P3Y, B00P59, B00P5P, B00PB4, B00PJJ, B00PPP, B00PQ7, and P002A20090500005.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- USA, Australia, Austria, Belgium, China, France, Germany, Hong Kong, India, Korea, Netherlands, Singapore, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom.
  • Product Description
    PercuNav Field Generator Stand is a Computed tomography, X-Ray System. Part Number 453561431911. || Designed to be used only with the PercuNav" image guided intervention system. The Field Generator Stand (with articulating Mounting Arm) is used to position the Field Generator near the area of interest.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips Medical Systems North America Co. Phillips, 22100 Bothell Everett Hwy, Bothell WA 98021-8431
  • Source
    USFDA