Recall of OrthoCase software

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Merge Healthcare, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    75909
  • Event Risk Class
    Class 2
  • Event Number
    Z-0878-2017
  • Event Initiated Date
    2015-10-08
  • Event Date Posted
    2016-12-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2018-06-22
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, image processing, radiological - Product Code LLZ
  • Reason
    Measurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment.
  • Action
    Merge issued a letter dated March 28, 2016, via email on April 4, 2016, or via certified mail if they did not have an email address for the customer. The letter notifies the customer of the issue, informs them of a workaround and that a fix is not yet available but they will be notified once one is available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. The response was required no later than April 18, 2016. Nonresponders were issued a second letter dated July 11, 2016, asking for a response by August 2, 2016. For further questions, please call (877) 741-5369.

Device

  • Model / Serial
    Versions: 2.0, 2.1, 2.1.1, 3.1, 3.2, 3.2.4, 3.3, 3.3.1, 3.4, 3.5.1, 3.5.1.1, 3.5.2, 3.5.3, 3.5.4, 3.5.5, 3.6, 3.6.1, 3.6.2, 3.7, 3.7.1, 3.7.2, 3.7.2 Vet, and 3.7.3
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distribution was made to medical facilities in CA, CO, CT, MA, MD, MN, MT, NE, NV, NY NJ, PA, SC, WA, and WI. Foreign distribution was made to Canada, as well as other countries.
  • Product Description
    Merge OrthoCase software. The firm name on the label is Merge Healthcare, Hartland, WI.
  • Manufacturer

Manufacturer