Recall of Orbit 90

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by ICU Medical, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    38209
  • Event Risk Class
    Class 2
  • Event Number
    Z-1166-2007
  • Event Date Posted
    2007-08-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-03-22
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    infusion set - Product Code FPA
  • Reason
    The product contains a female luer which has been reported to crack in use allowing fluid to leak from the device.
  • Action
    Firm mailed a letter dated September 26, 2006 to notify customers of the recall. The letter advises the product contains a female luer which has been reported to crack in use allowing fluid to leak from the device. Customers are asked that if they have inventory of the lots identified to quarantine these lots and contact your consignees of this recall and return the product immediately for destruction. They are asked to complete and return the attached product inventory summary sheet in the enclosed self-addressed stamped envelope and return product to ICU Medical.

Device

  • Model / Serial
    414371,414372,414373,433069,433070,453469,453470,453471,488549,488550,488551,488552,488553,488554,
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    VA, OH, TX, CA, GA
  • Product Description
    Orbit 90 Subcutaneous Infusion Set, 9 mm Catheter, 42" Tubing - Blue, List #P4291