Recall of NeuViz 16 MultiSliced CT Scanner System and NeuViz Dual series CT Scanner System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips And Neusoft Medical Systems Co., Ltd..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    63553
  • Event Risk Class
    Class 2
  • Event Number
    Z-0466-2013
  • Event Initiated Date
    2012-10-18
  • Event Date Posted
    2012-12-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2013-07-08
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, x-ray, tomography, computed - Product Code JAK
  • Reason
    The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. this issue may affect neuviz dual and neuviz 16 scanners.
  • Action
    Philips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to: 1. Have a mechanic check the patient table top screws fixing. Detailed instructions are attached as Annex with the letter. 2. If it is found that table top screws are improperly assembled, stop using the product and contact local Service personnel for repair. Philips and Neusoft Medical Systems Co., Ltd is in the process of upgrading devices with aforementioned kits and will contact customers to arrange for an update to affected systems. If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems Co. Ltd. at nms-service@neusoft.com For questions regarding this recall call 425-487-7665.

Device

  • Model / Serial
    NeuViz 16 system Serial numbers:  N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024.   Neuviz Dual Serial numbers:  400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
  • Product Description
    NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 || The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips And Neusoft Medical Systems Co., Ltd., 16 Century Road, Neusoft Park,, Hun Nan Industrial Area, Shenyang, Liaoning China
  • Source
    USFDA