Recall of Nanosphere

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Nanosphere, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69034
  • Event Risk Class
    Class 2
  • Event Number
    Z-2683-2014
  • Event Initiated Date
    2013-10-10
  • Event Date Posted
    2014-09-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-06-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Gram-positive bacteria and their resistance markers - Product Code PAM
  • Reason
    Nanosphere has received several reports relating to an increased rate of processing errors during procedures performed on the processor sp caused by tips in tip holder assemblies for the following products and lots of extraction trays.
  • Action
    Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.

Device

  • Model / Serial
    Product Number: 20-009-018; Lot Numbers and Expiration Dates: 1) Lot: 090613018B, Expiration Date: 03/05/14; 2) Lot: 091413018B, Expiration Date: 03/13/14; 3) Lot: 091713018C, Expiration Date: 03/16/14; 4) Lot: 091813018B, Expiration Date: 03/17/14; 5) Lot: 092113018C, Expiration Date: 03/20/14; 6) Lot: 100213018B, Expiration Date: 03/31/14
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
  • Product Description
    The Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-UP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-UP is performed directly on blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria. BC-UP detects and identifies the following bacterial genera and species: Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Enterococcus faecalis, Enterococcus faecium, Streptococcus spp., Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus anginosus, group and Listeria spp. || In addition, BC-UP detects the mnecA resistance marker, inferring mecA-mediated, methicillin resistance, and the vanA and vanB resistance markers, inferring vanA/vanB-mediated vancomycin resistance. In mixed growth, BC-UP does not specifically attribute van-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis. || BC-UP is indicated for use in conjunction with other clinical and laboratory findings to, aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to, monitor these infections. Sub-culturing of positive blood cultures is necessary to recover, organisms for susceptibility testing, identification of organisms not detected by BC-UP, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.