Recall of MPM, CAM01 PAC 1

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Integra LifeSciences Corp.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    60836
  • Event Risk Class
    Class 3
  • Event Number
    Z-0944-2012
  • Event Initiated Date
    2011-10-20
  • Event Date Posted
    2012-02-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2013-06-18
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Device, monitoring, intracranial pressure - Product Code GWM
  • Reason
    Some of the connector cables associated with integra's intracerebral pressure (icp) monitors may cause temporary interference with the icp waveform on the monitor display.
  • Action
    Integra sent a Urgent Product Recall Notification letter dated October 20, 2011, via FedEx, email, and fax to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to examine their inventory at your earliest convenience. If you have an effected PAC1 cable, please complete the attached Recall Acknowledgement Form and return it to Integra Customer Service per the instructions on the form. An Integra Sales Representative will contact you to assist in the return and replacement of new cables at no cost. In the meanwhile, you may continue to use the cables until you have been contacted and replacement product has been provided. Distortion of the ICP wave form can be stopped or prevented by keeping the PACI catheter receptacle out of human contact. Please contact Integra Customer Service at 1-800-654-2873.

Device

  • Model / Serial
    MPM, CAMO1, PAC1 :  MEF1100103;MED1102303; MEB1101602; MEF1100203; MEC1100602;MED1101003; MED1102203; MEE1102003; MED1102103; MEC1102302; MEB1101402; MEE1101003; MEF1100703; MEE1101903; MEF1101103; MEB1100803; MEC1103002; MEE1100903; MED1101703; MED1101203; MEC1102902; MEB1101103; MEC1102002; MEB1101502; MED1102203; MEF1100603; MEE1101203; MEC1100102; MEE1101503; MEE1100703; MEF1100403; MED1101303; MED1101403; MEF1100303; MEC1100702; MED1101103; MEC1102402; MEC1101602; MEF1100903; MEE1101303; MED1102003; MEC1103302; MEE1101803; MEE1100803; MEF1100503; MEC1102502; MEC1101702; MEE1101403; MEC1103402; MEF1100803; MEC1102802; MEC1101802;MEF1101003; MEC1103102; MEC1101902; MEC1102102
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution-- USA (nationwide ) .
  • Product Description
    Integra PAC1 Mod 2 Cables used with the Integra MPM, CAM01 Monitors