Recall of MISYS

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Misys Healthcare Systems.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    33732
  • Event Risk Class
    Class 2
  • Event Number
    Z-0088-06
  • Event Date Posted
    2005-10-22
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2006-02-10
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Calculator/Data Processing Module, For Clinical Use - Product Code JQP
  • Reason
    Customer complaint during testing of module prior to production use. software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000, the system willfile the number 100 instead of number 100,000.
  • Action
    Telephone calls and emails dated 9/29/05 to all consignees in production. Product Safety Noticed dated 9/29/2005 was faxed to all consignees testing the product.

Device

Manufacturer

  • Manufacturer Address
    Misys Healthcare Systems, 4801 E Broadway Blvd, Tucson AZ 85711-3609
  • Source
    USFDA