Recall of Linear Straight Broach Handle, (Rasp Handle) EZ Clean" Linear Broach Handle

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Greatbatch Medical.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    77120
  • Event Risk Class
    Class 2
  • Event Number
    Z-2055-2017
  • Event Initiated Date
    2017-04-23
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Broach - Product Code HTQ
  • Reason
    Greatbatch medical has voluntarily initiated a recall for the linear straight broach handle (rasp handle) and ez clean" linear broach handle. greatbatch medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.
  • Action
    Greatbatch sent an Urgent Medical Device Recall letter dated April 23, 2017. The letter described the product being recalled and described Reason For Recall, Risk To Health, Replacement Devices, & Instructions to DJO Surgical Inc. Advised consignee to quarantine and return the affected product. If the product was further distributed they should contact the end users and share the recall information with them. Requested consignees to complete and return the "Field Recall Response Form" to Stericycle at Greatbatch4053@stericycle.com or fax it to 877-479-8063. For questions contact your Greatbatch Medical representative, or contact us at 1-763-951-8235, or e-mail us at fieldaction.center@integer.net. For further questions, please call (763) 951-8184.

Device

  • Model / Serial
    "="" v-b-tooltip=""> Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697. EZ Clean" Linear Broach Handle: 3508491.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    US Distribution to the state of : TX
  • Product Description
    Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Greatbatch Medical, 2300 Berkshire Ln N, Plymouth MN 55441-4575
  • Manufacturer Parent Company (2017)
  • Source
    USFDA