Recall of Jostra

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by MAQUET Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36252
  • Event Risk Class
    Class 2
  • Event Number
    Z-0158-2007
  • Event Initiated Date
    2006-09-01
  • Event Date Posted
    2006-11-07
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-12-19
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Heart Lung Perfusion System - Product Code DTQ
  • Reason
    The hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.
  • Action
    The firm sent a safety alert letter on 9/13/06 to all US customers by Federal Express Return Receipt.

Device

  • Model / Serial
    Article # 923391 The affected HL-20 base units include S/Ns 100001 through 100240. Hand cranks shipped between the dates of 8/15/05 to 8/14/06 have the potential to be defective.  Serial #s100155, 1025, 1026, 1032, 1331, 1332, 100011, 100015, 100016, 100017, 1272, 1417, 1480, 1500, 1653, 14031054, 14031055, 14031059, 14041074, 14041874, 100012, 100013, J14041874, 1584, 100108, 100109, 14041074, JOSTRA, 100009, 100010, 100130, 100131, 1625, 1626, 1627, 1628, 1629, 100135, 100136, 100137, 14031040, 14031068, 100188, 1019, 1020, 1021, 1295, 100021, 100022, 100116, 14031051, 14031052, 1507, 1508, 1509, 100017, 100219, 100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    worldwide
  • Product Description
    Jostra - HL-20 Heart Lung Integrated Perfusion System, Roller Pump Hand crank
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MAQUET Inc., 1140 US Highway 22, Bridgewater NJ 08807-2958
  • Source
    USFDA