Recall of Integra Titan Reverse Shoulder System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Integra LifeSciences Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69596
  • Event Risk Class
    Class 2
  • Event Number
    Z-0197-2015
  • Event Initiated Date
    2014-10-24
  • Event Date Posted
    2014-11-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-07-21
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, shoulder, semi-constrained, metal/polymer cemented - Product Code KWS
  • Reason
    A single lot of left and a single lot of right reverse shoulder system cutting templates were manufactured incorrectly. specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
  • Action
    The recalling firm will notify their consignees via letter delivered by traceable courier service. The consignee will be asked to complete and return a form to the recalling firm stating if they have affected product or not.

Device

  • Model / Serial
    Lot Numbers PM0278 and PM0277
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
  • Product Description
    Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. || The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.
  • Manufacturer

Manufacturer