Recall of Human lgG4 Subclass Liquid Reagent Kits For use on the Roche Cobas"6000 Product Code LK009.CB

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by The Binding Site Group, Ltd..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    76041
  • Event Risk Class
    Class 2
  • Event Number
    Z-1069-2017
  • Event Initiated Date
    2014-09-11
  • Event Date Posted
    2017-01-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-05-03
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Method, nephelometric, immunoglobulins (g, a, m) - Product Code CFN
  • Reason
    Customer complaints identified that samples containing extremely high igg4 levels exhibit antigen excess and will yield spuriously low igg4 results.
  • Action
    Binding Site sent an Important Information letter dated September 30, 2014, Customers were instructed to retain the notification for records and additional instructions with the firm stating they were updating the instructions for use. Customers with questions were advised to contact their Binding Site representative or technical.service@bindingsite.co.uk.

Device

  • Model / Serial
    All lots.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
  • Product Description
    Human lgG4 Subclass Liquid Reagent Kits For use on the || Roche Cobas"6000 || This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents || Product Code LK009.CB
  • Manufacturer

Manufacturer

  • Manufacturer Address
    The Binding Site Group, Ltd., 8 Calthorpe Road, Birmingham United Kingdom
  • Manufacturer Parent Company (2017)
  • Source
    USFDA