Recall of High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Toshiba American Med Sys Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    33694
  • Event Risk Class
    Class 2
  • Event Number
    Z-0198-06
  • Event Initiated Date
    2005-09-21
  • Event Date Posted
    2005-11-24
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-07-12
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, X-Ray, Fluoroscopic, Image-Intensified - Product Code JAA
  • Reason
    It was found during a study, arcing might occur in the transformer of the high voltage generator, xtp-8100g and xtbp-8100g displaying an error message and interrupting fluoroscopy.
  • Action
    Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.

Device

  • Model / Serial
    model numbers: XTP-8100G AND XTBP-8100G
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX
  • Product Description
    High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Toshiba American Med Sys Inc, 2441 Michelle Dr, Tustin CA 92781
  • Source
    USFDA