Recall of HD11

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips Medical Systems.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    37888
  • Event Risk Class
    Class 2
  • Event Number
    Z-0987-2007
  • Event Initiated Date
    2007-04-25
  • Event Date Posted
    2007-07-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-01-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Diagnostic Ultrasound System - Product Code IYN
  • Reason
    Potential for erroneous measurements which may lead to incorrect diagnosis based upon the presence of an inaccurate, undetectable error in clinical measurement or calculation. if the user exits an imaging mode while active annotation is in use, the system can enter an undefined state where the image scale calibration may not match the scale of the on-screen measurements and graphics and possib.
  • Action
    On 04/25/2007 firm sent letters dated 04/23/2007 to consignees informing them of the issue and that a new software release to correct the issue became available 04/19/2007. The firm's Field Service Representatives will contact consignees to install the new software at no charge to the consignee.

Device

  • Model / Serial
    DOMESTIC SERIAL NUMBERS:   US10670272, US10670282, US10670288, US10670296, US10670304, US10670314, US10670315, US10670316, US10670317, US10670319, US10771576, US10771580, US10771596, US10771610, US10771620, US10771621, US10771622, US10771623, US10771630, US10771631, US10771638, US10771640, US10771641, US10771647, US10771651, US10771658, US10771660, US10771661, US20530061, US20530091, US20670366, US20670392, US20670393, US20771672, US20771674, US20771675, US20771683, US20771685, US20771686, US20771688, US20771689, US20771696, US20771697, US20771718, US20771722, US20771725, US20771726, US20771729, US20771733, US20771734, US20771735, US20771738, US20771739, US20771749, US30670410, US30670411, US30670501, US30670502, US30670509, US30670521, US30771757, US30771757, US30771766, US30771771, US30771772, US30771774, US30771775, US30771782, US30771790, US30771793, US30771809, US30771827, US30771830, US30771831, US30771835, US30771847, US30771848, US30771858, US30771862, US30771865, US40670540, US40670541, US40670573, US40670600, US40670624, US50670631, US50670657, US50670673, US50670707, US50670737, US60530418, US60670757, US60670777, US60670840, US60670841, US60670861, US60670862, US60670872, US60670873, US60670879, US60670880, US70670906, US70670912, US70670913, US70670914, US70670915, US70670947, US70670953, US70670955, US70670975, US80670986, US80670987, US80670990, US80670991, US80670992, US80670997, US80671008, US80671039, US80671040, US80671079, US80671083, US80671084, US90404706, US90531014, US90531073, US90671117, US90671154, US90671161, US90671165, US90671170, US90671178, USD0531367, USD0570096, USD0570114, USD0570161, USD0570195, USD0570200, USD0570207, USD0671475, USD0671490, USD0671493, USD0671497, USD0671498, USD0671503, USD0671504, USD0671505, USD0671510, USD0671511, USD0671527, USD0671531, USD0671532, USD0671533, USD0671542, USD0671546, USD0671547, USD0671548, USD0671550, USD0671553, USD0671554, USD0671555, USD0671556, USD0671567, USD0671572, USN0531226, USN0531237, USN0531243, USN0531256, USN0570005, USN0570014, USN0570031, USN0570036, USN0570044, USN0671319, USN0671320, USN0671321, USN0671322, USN0671323, USN0671325, USN0671326, USN0671341, USN0671351, USN0671352, USN0671353, USN0671360, USN0671362, USN0671367, USN0671370, USN0671378, USN0671406, USN0671407, USN0671408, USN0671410, USN0671413, USN0671414, USN0671418, USN0671422, USN0671423, USO0531081, USO0531087, USO0671212, USO0671214, USO0671254, USO0671260, USO0671263, USO0671271, USO0671285, USO0671286, USO0671287, USO0671288, USO0671289, USO0671290, USO0671291, USO0671293, USO0671294, USO0671304, USO0671311, USO0671315, USO0671316, and USO0671317.    INTERNATIONAL SERIAL NUMBERS:  US10670302, US10732432, US10771574, US10771578, US10771588, US10771590, US10771591, US10771593, US10771594, US10771595, US10771597, US10771599, US10771601, US10771603, US10771606, US10771607, US10771608, US10771612, US10771617, US10771628, US10771629, US10771632, US10771633, US10771635, US10771636, US10771639, US10771644, US10771646, US10771649, US10771652, US10771654, US10771655, US10771656, US10771657, US10771662, US10771666, US10771667, US10771669, US10771671, US20670326, US20771673, US20771678, US20771680, US20771681, US20771687, US20771691, US20771695, US20771699, US20771701, US20771708, US20771713, US20771714, US20771715, US20771716, US20771727, US20771728, US20771744, US20771747, US30670469, US30670525, US30771756, US30771756, US30771758, US30771759, US30771768, US30771778, US30771780, US30771786, US30771788, US30771792, US30771797, US30771800, US30771802, US30771805, US30771807, US30771810, US30771811, US30771812, US30771815, US30771816, US30771818, US30771819, US30771820, US30771828, US30771832, US30771833, US30771836, US30771837, US30771840, US30771842, US30771843, US30771845, US30771846, US30771849, US30771853, US30771861, US30771861, US40631672, US40670548, US40670551, US40670575, US40670576, US40670577, US50530353, US50670674, US50670745, US60670785, US60670812, US60670813, US60670888, US70670905, US70670911, US70670920, US80530793, US80530809, US80671003, US80671014, US80671074, US80671082, US90671125, US90671175, US90671176, US90671183, US90671192, US90671195, USD0570097, USD0570116, USD0570121, USD0570149, USD0570154, USD0570155, USD0570166, USD0570169, USD0570171, USD0671440, USD0671441, USD0671442, USD0671443, USD0671444, USD0671445, USD0671446, USD0671447, USD0671448, USD0671449, USD0671450, USD0671451, USD0671452, USD0671453, USD0671454, USD0671455, USD0671457, USD0671458, USD0671459, USD0671460, USD0671461, USD0671462, USD0671463, USD0671464, USD0671465, USD0671466, USD0671467, USD0671468, USD0671469, USD0671470, USD0671471, USD0671472, USD0671473, USD0671474, USD0671476, USD0671477, USD0671478, USD0671479, USD0671480, USD0671481, USD0671482, USD0671483, USD0671484, USD0671485, USD0671486, USD0671487, USD0671488, USD0671489, USD0671491, USD0671492, USD0671494, USD0671495, USD0671496, USD0671499, USD0671500, USD0671501, USD0671502, USD0671506, USD0671507, USD0671508, USD0671509, USD0671512, USD0671513, USD0671514, USD0671515, USD0671516, USD0671517, USD0671518, USD0671519, USD0671520, USD0671521, USD0671522, USD0671523, USD0671524, USD0671525, USD0671534, USD0671537, USD0671538, USD0671543, USD0671544, USD0671545, USD0671562, USD0671563, USD0671566, USN0570007, USN0570008, USN0570009, USN0570055, USN0570056, USN0570057, USN0671324, USN0671327, USN0671328, USN0671329, USN0671330, USN0671331, USN0671332, USN0671333, USN0671334, USN0671335, USN0671336, USN0671337, USN0671338, USN0671339, USN0671340, USN0671342, USN0671343, USN0671344, USN0671345, USN0671346, USN0671347, USN0671348, USN0671349, USN0671354, USN0671355, USN0671356, USN0671357, USN0671358, USN0671359, USN0671361, USN0671363, USN0671364, USN0671365, USN0671366, USN0671368, USN0671369, USN0671371, USN0671372, USN0671373, USN0671374, USN0671375, USN0671376, USN0671379, USN0671380, USN0671381, USN0671382, USN0671383, USN0671384, USN0671385, USN0671386, USN0671387, USN0671388, USN0671389, USN0671390, USN0671391, USN0671392, USN0671393, USN0671394, USN0671395, USN0671396, USN0671397, USN0671398, USN0671399, USN0671400, USN0671401, USN0671402, USN0671403, USN0671404, USN0671405, USN0671409, USN0671411, USN0671412, USN0671415, USN0671416, USN0671417, USN0671419, USN0671420, USN0671421, USN0671424, USN0671426, USN0671427, USN0671428, USN0671429, USN0671430, USN0671431, USN0671432, USN0671433, USN0671434, USN0671435, USN0671436, USN0671437, USN0671438, USN0671439, USO0531134, USO0531137, USO0671213, USO0671215, USO0671279, USO0671280, USO0671281, USO0671282, USO0671283, USO0671284, USO0671292, USO0671295, USO0671296, USO0671297, USO0671298, USO0671299, USO0671300, USO0671301, USO0671302, USO0671303, USO0671305, USO0671306, USO0671307, USO0671308, USO0671309, USO0671310, USO0671312, USO0671313, USO0671314, and USO0671318.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide.
  • Product Description
    HD11XE Diagnostic Ultrasound System
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips Medical Systems, 22100 Bothell-Everett Hwy., P.O. Box 3003, Bothell WA 98041-3003
  • Source
    USFDA