Recall of Havard Clinical Technology

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Harvard Clinical Technology.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26903
  • Event Risk Class
    Class 2
  • Event Number
    Z-1208-03
  • Event Initiated Date
    2003-08-11
  • Event Date Posted
    2003-09-11
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-05-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Pump, Infusion - Product Code FRN
  • Reason
    Under specific sequence conditions, the unit may apper to be infusing, but is not.
  • Action
    Harvard Medical Technology inc. notified customers by letter on 8/11/03, advising users of the software upgrade to version V1.3T. Facilities will be visited by technicians who will upgrade the software.

Device

Manufacturer

  • Manufacturer Address
    Harvard Clinical Technology, 22 Pleasant Street, South Natick MA 01760
  • Source
    USFDA