Recall of GEHC OEC Injection Cable for MedRad Provis V Contrast Injector.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE OEC Medical Systems, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62746
  • Event Risk Class
    Class 2
  • Event Number
    Z-2262-2012
  • Event Initiated Date
    2008-10-10
  • Event Date Posted
    2012-08-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2013-01-10
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
  • Reason
    Ge initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
  • Action
    An Urgent Safety Advisory Notice was sent to consignees on October 10, 2008. The letter identified the affected product and notified customers of the situation, along with instructing them to discontinue use until proper validation and instruction update for cable use was completed. GEHC OEC was to have contacted customers when the solution became available and resolved it at no cost to the customer.

Device

  • Model / Serial
    Serial Numbers: 82-1138, 82-1586, 82-2089, 82-2117, 82-2182, 82-2214, 82-2226, 82-2262, 82-2264, 82-2308, 82-2312, 82-2314, 82-2321, 82-2328, 82-2380, 82-2391, 82-2438, 82-2483, 82-2512, 82-2516, 82-2535, 82-2722, 82-2796, 82-2821, 82-2846, 82-2848, 82-3052, 82-3055, 82-3129, 82-3144, 82-3170, 82-3171, 82-3201, 82-3256, 82-3267, 82-3314, 82-3350, 82-3361, 82-3377, 82-3467, 82-3468, 82-3659, 82-3709, 82-3722, 82-3859, 82-7016-MH, 82-7022-MH, 82-7037-MH, 82-7055-MH, 82-7062-MH, 82-7080-MH, 82-7086-MH, 82-7105-MH, 82-7114-MH, 82-7124-MH, 82-7131-MH, 82-7133-MH, 82-7137-MH, 82-7141-MH, 82-7154-CMH, 82-7155-MH, 82-7156-MH, 82-7161-MH, 82-7175-MH, 82-7189-MH, 82-7190-MH, 89-1441, 8S-1053, 8S-1436, 8S-2059-C, 8S-7021-CMH, 8S-7023-CMH, 8S-7040-CMH, 8S-7087-CMH, 8S-7164-CMH, DEMO/8S-1355-N, DEMO/E3-7002-MH, E2-0049, E2-0088, E2-0111, E2-0262, E2-7001-MH, E2-7007-MH, E2-7010-MH, E2-7014-MH, E2-7015-MH, E2-7031-MH, E2-7032-MH, E2-7035-MH, E2-7043-MH, E2-7059-MH, E2-7061-MH, E2-7067-MH, E2-7075-MH, E2-7076-MH, E2-7015-MH, & rENTAL/82-7004-MH.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide distribution.
  • Product Description
    GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. || In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger a contrast injection. Part numbers include 00-902260-01 and 00-883654-01.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE OEC Medical Systems, Inc, 384 Wright Brothers Dr, Salt Lake City UT 84116-2862
  • Source
    USFDA