Recall of GE Healthcare Inc.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE Healthcare.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    71964
  • Event Risk Class
    Class 2
  • Event Number
    Z-2470-2015
  • Event Initiated Date
    2015-08-11
  • Event Date Posted
    2015-08-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-01-05
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, image processing, radiological - Product Code LLZ
  • Reason
    Images from the centricity pacs-iw with universal viewer and centricity universal viewer may be missing when a system parameter maproute is set to a value greater than 1.
  • Action
    The firm sent out an Urgent Medical Device Correction letter dated 8/11/2015 to all affected customers. Letter states that No User activity required. GE Healthcare has completed a remote inspection of your system and found that this system parameter is currently set to the default setting of 1. The firm will correct all affected systems by a providing software update. The letter also states that if the customer has any questions they should call GE Healthcare Service at 1-800-437-1171 or their local service representative.

Device

  • Model / Serial
    Model Number: Centricity Universal Viewer Version 6.0 software  Codes: 2090255-001 UV 6.0 Web Client Media Kit. 05 Jun 15 to present
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - U.S. nationwide in the states of: AK, CA, CO, DE, FL, HI, ID, NY, PA, SD, TN and TX; France, Germany, Greece, India, Israel, Malaysia, Mexico, Northern Ireland (UK), Panama, Poland, Russia, South Africa, Saudi Arabia, Switzerland, Turkey, UAE and Uruguay.
  • Product Description
    Centricity Universal Viewer || Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving and communications system, a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. Used in Radiology.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE Healthcare, 540 W Northwest Hwy, Barrington IL 60010-3051
  • Manufacturer Parent Company (2017)
  • Source
    USFDA