Recall of GE Healthcare DatexOhmeda Aisys

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Datex - Ohmeda, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36057
  • Event Risk Class
    Class 2
  • Event Number
    Z-0021-2007
  • Event Initiated Date
    2006-08-10
  • Event Date Posted
    2006-11-07
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-02-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Anesthesia Delivery System - Product Code BSZ
  • Reason
    Vaporizer shutdown with alarm-an intermittent leak in the aladin 2 cassette/aisys pneumatic interface (valves) may occur that will invoke an alarm state. when this occurs 'vaporizer failure' will appear on the aisys main display, with an accompanying audible alrm. delivery of anesthetic agent is halted; gas flow and ventilation are not affected.
  • Action
    Firm mailed recall letters on August 10, 2006, to consignees telling them what to do if a failure occurred during system checkout or during use, as a short term solution. Later, GE representatives will make visits to each site for modification of the interface valve on each device, as the long term solution.

Device

  • Model / Serial
    Serial numbers: ANAK00115, ANAH00114, ANAH00115, ANAH00116, ANAH00117, ANAH00124, ANAH00126, ANAJ00100, ANAJ00101, ANAJ00102, ANAJ00103, ANAJ00104, ANAJ00105, ANAJ00106, ANAJ00107, ANAJ00108, ANAJ00109, ANAJ00110, ANAJ00111, ANAJ00112, ANAJ00113, ANAJ00114, ANAJ00115, ANAJ00116, ANAJ00117, ANAJ00118, ANAJ00119, ANAJ00120, ANAJ00121, ANAJ00122, ANAJ00123, ANAJ00124, ANAJ00125, ANAJ00126, ANAJ00127, ANAJ00128, ANAJ00129, ANAJ00131, ANAJ00132, ANAJ00133, ANAJ00134, ANAJ00135, ANAJ00136, ANAJ00137, ANAJ00138, ANAJ00139, ANAJ00140, ANAJ00141, ANAJ00142, ANAJ00143, ANAJ00144, ANAJ00145, ANAJ00146, ANAJ00147, ANAJ00148, ANAJ00149, ANAJ00150, ANAJ00151, ANAJ00152, ANAJ00153, ANAJ00154, ANAJ00155, ANAJ00156, ANAJ00157, ANAJ00158, ANAJ00159, ANAJ00160, ANAJ00161, ANAJ00162, ANAJ00163, ANAJ00164, ANAJ00165, ANAJ00166, ANAJ00167, ANAJ00168, ANAJ00169, ANAJ00170, ANAJ00171, ANAJ00172, ANAJ00173, ANAJ00174, ANAJ00175, ANAJ00176, ANAJ00177, ANAJ00178, ANAJ00179, ANAJ00180, ANAJ00181, ANAJ00182, ANAJ00183, ANAJ00184, ANAJ00185, ANAJ00186, ANAJ00187, ANAJ00188, ANAJ00189, ANAJ00190, ANAJ00191, ANAJ00192, ANAJ00193, ANAJ00194, ANAJ00195, ANAJ00195, ANAJ00196, ANAJ00197, ANAJ00198, ANAJ00199, ANAJ00200, ANAJ00201, ANAJ00202, ANAJ00203, ANAJ00204, ANAJ00205, ANAJ00206, ANAJ00207, ANAJ00208, ANAJ00209, ANAJ00210, ANAJ00211, ANAJ00212, ANAJ00213, ANAJ00214, ANAJ00215, ANAJ00216, ANAJ00217, ANAJ00218, ANAJ00219, ANAJ00220, ANAJ00221, ANAJ00222, ANAJ00223, ANAJ00224, ANAJ00225, ANAJ00226, ANAJ00227, ANAJ00228, ANAJ00229, ANAJ00230, ANAJ00231, ANAJ00232, ANAJ00233, ANAJ00234, ANAJ00235, ANAJ00236, ANAJ00237, ANAJ00238, ANAJ00239, ANAJ00240, ANAJ00241, ANAJ00242, ANAJ00243, ANAJ00244, ANAJ00245, ANAJ00246, ANAJ00247, ANAJ00248, ANAJ00249, ANAJ00250, ANAJ00251, ANAJ00252, ANAJ00253, ANAJ00254, ANAJ00255, ANAJ00256, ANAJ00257, ANAJ00258, ANAJ00259, ANAJ00260, ANAJ00261, ANAJ00262, ANAJ00263, ANAJ00264, ANAJ00265, ANAJ00266, ANAJ00267, ANAK00100, ANAK00101, ANAK00102, ANAK00103, ANAK00104, ANAK00105, ANAK00106, ANAK00107, ANAK00108, ANAK00109, ANAK00110, ANAK00111, ANAK00112, ANAK00113, ANAK00114, ANAK00116, ANAK00117, ANAK00118, ANAK00119, ANAK00120, ANAK00121, ANAK00122, ANAK00123, ANAK00124, ANAK00125, ANAK00126, ANAK00127, ANAK00128, ANAK00129, ANAK00130, ANAK00131, ANAK00132, ANAK00133, ANAK00134, ANAK00135, ANAK00136, ANAK00137, ANAK00138, ANAK00139, ANAK00140, ANAK00141, ANAK00142, ANAK00143, ANAK00144, ANAK00145, ANAK00146, ANAK00147, ANAK00148, ANAK00149, ANAK00150, ANAK00151, ANAK00152, ANAK00153, ANAK00154, ANAK00155, ANAK00156, ANAK00156, ANAK00157, ANAK00158, ANAK00159, ANAK00160, ANAK00161, ANAK00162, ANAK00163, ANAK00164, ANAK00165, ANAK00166, ANAK00167, ANAK00168, ANAK00169, ANAK00170, ANAK00171, ANAK00172, ANAK00173, ANAK00174, ANAK00175, ANAK00176, ANAK00177, ANAK00178, ANAK00179, ANAK00180, ANAK00181, ANAK00182, ANAK00183, ANAK00184, ANAK00185, ANAK00186, ANAK00187, ANAK00188, ANAK00189, ANAK00190, ANAK00191, ANAK00192, ANAK00193, ANAK00194, ANAK00195, ANAK00196, ANAK00197, ANAK00198, ANAK00199, ANAK00200, ANAK00201, ANAK00202, ANAK00203, ANAK00204, ANAK00205, ANAK00206, ANAK00207, ANAK00208, ANAK00209, ANAK00210, ANAK00211, ANAK00212, ANAK00213, ANAK00214, ANAK00215, ANAK00216, ANAK00217, ANAK00218, ANAK00219, ANAK00220, ANAK00221, ANAK00222, ANAK00223, ANAK00224, ANAK00225, ANAK00226, ANAK00227, ANAK00228, ANAK00229, ANAK00230, ANAK00231, ANAK00232, ANAK00233, ANAK00234, ANAK00235, ANAK00236, ANAK00237, ANAK00238, ANAK00239, ANAK00240, ANAK00241, ANAK00242, ANAK00243, ANAK00244, ANAK00245, ANAK00246, ANAK00247, ANAK00248, ANAK00249, ANAK00250, ANAK00251, ANAK00252, ANAK00253, ANAK00254, ANAK00255, ANAK00256, ANAK00257, ANAK00258, ANAK00259, ANAK00260, ANAK00261, ANAK00262, ANAK00263, ANAK00264, ANAK00265, ANAK00266, ANAK00267, ANAK00268, ANAK00269, ANAK00270, ANAK00271, ANAK00272, ANAK00273, ANAK00274, ANAK00275, ANAK00276, ANAK00277, ANAK00278, ANAK00279, ANAK00280, ANAK00281, ANAK00282, ANAK00283, ANAK00284, ANAK00285, ANAK00286, ANAK00287, ANAK00288, ANAK00289, ANAK00290, ANAK00291, ANAK00292, ANAK00293, ANAK00294, ANAK00295, ANAK00296, ANAK00297, ANAK00298, ANAK00299, ANAK00300, ANAK00301, ANAK00302, ANAK00303, ANAK00304, ANAK00305, ANAK00306, ANAK00307, ANAK00308, ANAK00309, ANAK00310, ANAK00311, ANAK00312, ANAK00313, ANAK00314, ANAK00315, ANAK00316, ANAK00317, ANAK00318, ANAK00319, ANAK00320, ANAK00321, ANAK00322, ANAK00323, ANAK00324, ANAK00325, ANAK00326, ANAK00327, ANAK00329, ANAK00330, ANAK00331, ANAK00332, ANAK00333, ANAK00334, ANAK00335, ANAK00336, ANAK00337, ANAK00338, ANAK00339, ANAK00340, ANAK00341, ANAK00342, ANAK00343, ANAK00344, ANAK00348, ANAK00349, ANAK00350, ANAK00351, ANAK00352, ANAK00353, ANAK00354, ANAK00355, ANAK00356, ANAK00357, ANAK00358, ANAK00359, ANAK00360, ANAK00361, ANAK00362, ANAK00363, ANAK00364, ANAK00365, ANAK00366, ANAK00367, ANAK00368, ANAK00369, ANAK00370, ANAK00371, ANAK00372, ANAK00373, ANAK00374, ANAK00375, ANAK00376, ANAK00377, ANAK00378, ANAK00379, ANAK00380, ANAK00381, ANAK00382, ANAK00383, ANAK00384, ANAK00385, ANAK00386, ANAK00387, ANAK00388, ANAK00389, ANAK00390, ANAK00391, ANAK00392, ANAK00393, ANAK00394, ANAK00395, ANAK00412, ANAK00413, ANAK00414, ANAK00415,
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    AK, AL, CA, FL, KS, KY, MD, MN, MO, NC, NE, SC, SD, TX, VA, WI,
  • Product Description
    GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Datex - Ohmeda, Inc, Po Box 7550, Madison WI 53707
  • Source
    USFDA