Recall of Fluoro CArm Xray system.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE OEC Medical Systems, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36099
  • Event Risk Class
    Class 2
  • Event Number
    Z-1302-06
  • Event Initiated Date
    2006-08-16
  • Event Date Posted
    2006-09-30
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-07-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    X-ray - Product Code JAA
  • Reason
    X-ray systems could provide output which exceeds the 20 r/minute limit.
  • Action
    Consignees were notified by certified letter on 08/14-15/2006.

Device

  • Model / Serial
    All units
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide
  • Product Description
    GE OEC 9800 Fluoro C-Arm X-ray System, GE OEC Medical Systems Inc.