Recall of Expo Angiographic (Diagnostic) Catheter

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Boston Scientific Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    52826
  • Event Risk Class
    Class 2
  • Event Number
    Z-2015-2009
  • Event Initiated Date
    2009-08-05
  • Event Date Posted
    2009-09-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-01-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Catheter, intravascular, diagnostic - Product Code DQO
  • Reason
    Boston scientific is initiating a recall removal of numerous batches/lots of wiseguide guide catheter, impulse angiographic (diagnostic) catheter, expo angiographic (diagnostic) catheter. boston scientific has determined that the sterile barrier in the packaging may be compromised. during internal inspections, they discovered gaps between the side seals and the top seal approximately 3/8 in len.
  • Action
    Boston Scientific issued an "Urgent Medical Device Recall Removal" letter dated August 5, 2009. The letter was addressed to Risk Manger / Field Action Contact. The letter described the problem and product involved. The Firm recommended to segregate affected product immediately and return to Boston Scientific and requested return of the "The Reply Verification Tracking Form" enclosed with the letter. For further questions, contact your local Boston Scientific Sales Representative or call 1-763-494-1634.

Device

  • Model / Serial
    Material Description, Batch;  MODEL D 5F FR3.5 SINGLE 20165903;  MODEL D 5F FR3.5 SINGLE 20173017;  MODEL D 5F FR3.5 (5PK) 20155187;  MODEL D 5F FR3.5 (5PK) 20160744;  MODEL D 5F FR3.5 (5PK) 20162419;  MODEL D 5F FR3.5 (5PK) 20167751;  MODEL D 5F FR3.5 (5PK) 20168851;  MODEL D 5F FR3.5 (5PK) 20170546;  MODEL D 5F FR3.5 (5PK) 20174318;  MODEL D 5F FR3.5 (5PK) 20178095;  MODEL D 5F FR3.5 (5PK) 20179470;  MODEL D 5F FR4 SINGLE 20155421;  MODEL D 5F FR4 SINGLE 20162599;  MODEL D 5F FR4 SINGLE 20167927;  MODEL D 5F FR4 (5PK) 20155189;  MODEL D 5F FR4 (5PK) 20156776;  MODEL D 5F FR4 (5PK) 20156777;  MODEL D 5F FR4 (5PK) 20156778;  MODEL D 5F FR4 (5PK) 20158084;  MODEL D 5F FR4 (5PK) 20160746;  MODEL D 5F FR4 (5PK) 20162420;  MODEL D 5F FR4 (5PK) 20162421;  MODEL D 5F FR4 (5PK) 20164300;  MODEL D 5F FR4 (5PK) 20164301;  MODEL D 5F FR4 (5PK) 20165769;  MODEL D 5F FR4 (5PK) 20165770;  MODEL D 5F FR4 (5PK) 20165771;  MODEL D 5F FR4 (5PK) 20167752;  MODEL D 5F FR4 (5PK) 20167753;  MODEL D 5F FR4 (5PK) 20167754;  MODEL D 5F FR4 (5PK) 20168852;  MODEL D 5F FR4 (5PK) 20170547;  MODEL D 5F FR4 (5PK) 20172880;  MODEL D 5F FR4 (5PK) 20172881;  MODEL D 5F FR4 (5PK) 20174319;  MODEL D 5F FR4 (5PK) 20178096;  MODEL D 5F FR4 (5PK) 20178098;  MODEL D 5F FR4 (5PK) 20178099;  MODEL D 5F FR4 (5PK) 20178100;  MODEL D 5F FR4 (5PK) 20179471;  MODEL D 5F FR4 (5PK) 20179472;  MODEL D 5F FR4 (5PK) 20181248;  MODEL D 5F FR4 (5PK) 20181249;  MODEL D 5F FR4 (5PK) 20181250;  MODEL D 5F FR5 (5PK) 20155190;  MODEL D 5F FR5 (5PK) 20160747;  MODEL D 5F FR5 (5PK) 20162422;  MODEL D 5F FR5 (5PK) 20164302;  MODEL D 5F FR5 (5PK) 20165772;  MODEL D 5F FR5 (5PK) 20167755;  MODEL D 5F FR5 (5PK) 20174320;  MODEL D 5F FR6 (5PK) 20164303;  MODEL D 5F FR4 MODIFIED (5PK) 20160748;  MODEL D 5F AL3 (5PK) 20156779;  MODEL D 5F AL3 (5PK) 20179474;  MODEL D 5F MPA-1 SINGLE 20165904;  MODEL D 5F MPA-1 W/O SH (5PK) 20153941;  MODEL D 5F MPA-1 W/O SH (5PK) 20155191;  MODEL D 5F MPA-1 W/O SH (5PK) 20162423;  MODEL D 5F MPA-1 W/O SH (5PK) 20164304;  MODEL D 5F MPA-1 W/O SH (5PK) 20165773;  MODEL D 5F MPA-1 W/O SH (5PK) 20170548;  MODEL D 5F MPA-1 W/O SH (5PK) 20174322;  MODEL D 5F MPA-1 W/O SH (5PK) 20178102;  MODEL D 5F MPA2 SH SINGLE 20173019;  MODEL D 5F MPA2 W/SH (5PK) 20153942;  MODEL D 5F MPA2 W/SH (5PK) 20155192;  MODEL D 5F MPA2 W/SH (5PK) 20162424;  MODEL D 5F MPA2 W/SH (5PK) 20172883;  MODEL D 5F MPA2 W/SH (5PK) 20174323;  MODEL D 5F MPA2 W/SH (5PK) 20178103;  MODEL D 5F MPA2 W/SH (5PK) 20179476;  MODEL D 5F MPB-2 SH SINGLE 20156904;  MODEL D 5F MPB-2 SH SINGLE 20174504;  MODEL D 5F MPB-2 W/ SH (5PK) 20160750;  MODEL D 5F MPB-2 W/ SH (5PK) 20164305;  MODEL D 5F CAS2 (5PK) 20178104;  MODEL - EXPO 5F SELECTIVE KIMNY 5 PACK 20178105;  MODEL D 5F RCB SINGLE 20167929; MODEL D 5F RCB (5PK) 20155194;  MODEL D 5F RCB (5PK) 20172885;  MODEL D 5F RCB (5PK) 20174324;  MODEL D 5F RCB (5PK) 20178106;  MODEL D 5F RCB (5PK) 20179477;  MODEL D 5F RCB (5PK) 20181255;  MODEL D 5F LCB (5PK) 20162425;  MODEL D 5F LCB (5PK) 20167758;  MODEL D 5F LCB (5PK) 20172886;  MODEL D 5F LCB (5PK) 20174325;  MODEL D 5F LCB (5PK) 20178107;  MODEL D 5F IM SINGLE 20158218;  MODEL D 5F IM SINGLE 20178281;  MODEL D 5F IM (5PK) 20155195; MODEL D 5F IM (5PK) 20156780; MODEL D 5F IM (5PK) 20162426; MODEL D 5F IM (5PK) 20164306; MODEL D 5F IM (5PK) 20165774; MODEL D 5F IM (5PK) 20167759;  MODEL D 5F IM (5PK) 20170551;  MODEL D 5F IM (5PK) 20172887;  MODEL D 5F IM (5PK) 20174326;  MODEL D 5F IM (5PK) 20178108;  MODEL D 5F IM (5PK) 20181257;  MODEL - D 5F IM T (5PK) 20178109;  MODEL-EXPO 5F SEL IMC-CUMMINGS CURVE 5PK 20167761;  MODEL-EXPO 5F SEL IMC-CUMMINGS CURVE 5PK 20181261;  MODEL - 5F EXPO FL3 5-PK 20151688; MODEL D 5F FL3.5 SINGLE 20155422; MODEL D 5F FL3.5 SINGLE 20158219;  MODEL D 5F FL3.5 SINGLE 20164492;  MODEL D 5F FL3.5 (5PK) 12687931;  MODEL D 5F FL3.5 (5PK) 20155196;  MODEL D 5F FL3.5 (5PK) 20156781;  MODEL D 5F FL3.5 (5PK) 20158094;  MODEL D 5F FL3.5 (5PK) 20164307;  MODEL D 5F FL3.5 (5PK) 20165775;  MODEL D 5F FL3.5 (5PK) 20167762;  MODEL D 5F FL3.5 (5PK) 20168858;  MODEL D 5F FL3.5 (5PK) 20170552;  MODEL D 5F FL3.5 (5PK) 20172888;  MODEL D 5F FL3.5 (5PK) 20174327;  MODEL D 5F FL3.5 (5PK) 20178111;  MODEL D 5F FL3.5 (5PK) 20179479;  MODEL D 5F FL3.5 (5PK) 20181263;  MODEL D 5F FL4 SINGLE 20158220;  MODEL D 5F FL4 SINGLE 20160864;  MODEL D 5F FL4 (5PK) 20155197;  MODEL D 5F FL4 (5PK) 20156784;  MODEL D 5F FL4 (5PK) 20156785;  MODEL D 5F FL4 (5PK) 20158095;  MODEL D 5F FL4 (5PK) 20158096;  MODEL D 5F FL4 (5PK) 20160752;  MODEL D 5F FL4 (5PK) 20162427;  MODEL D 5F FL4 (5PK) 20162428;  MODEL D 5F FL4 (5PK) 20164310;  MODEL D 5F FL4 (5PK) 20164311;  MODEL D 5F FL4 (5PK) 20165776;  MODEL D 5F FL4 (5PK) 20165777;  MODEL D 5F FL4 (5PK) 20167765;  MODEL D 5F FL4 (5PK) 20167766;  MODEL D 5F FL4 (5PK) 20168859;  MODEL D 5F FL4 (5PK) 20170554;  MODEL D 5F FL4 (5PK) 20172889;  MODEL D 5F FL4 (5PK) 20172890;  MODEL D 5F FL4 (5PK) 20174329;  MODEL D 5F FL4 (5PK) 20178113;  MODEL D 5F FL4 (5PK) 20178114;  MODEL D 5F FL4 (5PK) 20179481;  MODEL D 5F FL4 (5PK) 20179482;  MODEL D 5F FL4 (5PK) 20179483;  MODEL D 5F FL4 (5PK) 20181264;  MODEL D 5F FL4.5 SINGLE 20162600;  MODEL D 5F FL4.5 (5PK) 20155198;  MODEL D 5F FL4.5 (5PK) 20162429;  MODEL D 5F FL4.5 (5PK) 20172892;  MODEL D 5F FL5 SINGLE 20164495;  MODEL D 5F FL5 SINGLE 20167932;  MODEL D 5F FL5 (5PK) 20155199;  MODEL D 5F FL5 (5PK) 20156786;  MODEL D 5F FL5 (5PK) 20162431;  MODEL D 5F FL5 (5PK) 20164312;  MODEL D 5F FL5 (5PK) 20165778;  MODEL D 5F FL5 (5PK) 20167769;  MODEL D 5F FL5 (5PK) 20170555;  MODEL D 5F FL5 (5PK) 20172894;  MODEL D 5F FL5 (5PK) 20174332;  MODEL D 5F FL5 (5PK) 20178116;  MODEL D 5F FL5 (5PK) 20179484;  MODEL D 5F FL5 (5PK) 20181266;  MODEL D 5F FL6 (5PK) 20155200;  MODEL D 5F FL6 (5PK) 20168862;  MODEL D 5F FL6 (5PK) 20179485;  MODEL D 5F PIG 110 SINGLE 20174508;  MODEL D 5F PIG 110 CM (5PK) 20158101;  MODEL D 5F PIG 110 CM (5PK) 20160757;  MODEL D 5F PIG 110 CM (5PK) 20162435;  MODEL D 5F PIG 110 CM (5PK) 20164316;  MODEL D 5F PIG 110 CM (5PK) 20168866;  MODEL D 5F PIG 110 CM (5PK) 20170559;  MODEL D 5F PIG 110 CM (5PK) 20172898;  MODEL D 5F PIG 110 CM (5PK) 20174336;  MODEL D 5F PIG 110 CM (5PK) 20178120;  MODEL D 5F PIG 110 CM (5PK) 20179489;  MODEL D 5F PIG 110 CM (5PK) 20181270;  MODEL D 5F PIG 145 110 SINGLE 20156909;  MODEL D 5F PIG 145 110 SINGLE 20167936;  MODEL D 5F PIG 145 110 SINGLE 20179659;  MODEL D 5F PIG 145 110 CM (5PK) 20158102;  MODEL D 5F PIG 145 110 CM (5PK) 20160758;  MODEL D 5F PIG 145 110 CM (5PK) 20162436;  MODEL D 5F PIG 145 110 CM (5PK) 20164317;  MODEL D 5F PIG 145 110 CM (5PK) 20168867;  MODEL D 5F PIG 145 110 CM (5PK) 20170560;  MODEL D 5F PIG 145 110 CM (5PK) 20172899;  MODEL D 5F PIG 145 110 CM (5PK) 20174337;  MODEL D 5F PIG 145 110 CM (5PK) 20178121;  MODEL D 5F PIG 145 110 CM (5PK) 20179490;  MODEL D 5F PIG 145 110 CM (5PK) 20179491;  MODEL D 5F PIG 145 110 CM (5PK) 20181271;  MODEL D 5F PIG 145 110 CM (5PK) 20181272;  MODEL D 5F PIG 155 110 CM (5PK) 20156792;  MODEL D 5F PIG 155 110 CM (5PK) 20160759;  MODEL D 5F PIG 155 110 CM (5PK) 20162437;  MODEL D 5F PIG 155 110 CM (5PK) 20168868;  MODEL D 5F PIG 155 110 CM (5PK) 20170561;  MODEL D 5F PIG 155 110 CM (5PK) 20172900;  MODEL D 5F PIG 155 110 CM (5PK) 20174338;  MODEL D 5F PIG 155 110 CM (5PK) 20178123;  MODEL D 5F PIG 155 110 CM (5PK) 20179492;  MODEL-5F EXPO CLS4 5-PK 20151687; MODEL D 5F AR1 (5PK) 20155209; MODEL D 5F AR1 (5PK) 20162438;  MODEL D 5F AR1 (5PK) 20164318;  MODEL D 5F AR1 (5PK) 20167778;  MODEL D 5F AR1 (5PK) 20172901;  MODEL D 5F AR1 (5PK) 20178124;  MODEL D 5F AR2 SINGLE 20178282;  MODEL D 5F AR2 (5PK) 20155210;  MODEL D 5F AR2 (5PK) 20156793;  MODEL D 5F AR2 (5PK) 20162439;  MODEL D 5F AR2 (5PK) 20167779;  MODEL D 5F AR2 (5PK) 20172902;  MODEL D 5F AR2 (5PK) 20174339;  MODEL D 5F AR MOD SINGLE 20165905;  MODEL D 5F AR MODIFIED (5PK) 20155211;  MODEL D 5F AR MODIFIED (5PK) 20155212;  MODEL D 5F AR MODIFIED (5PK) 20155213;  MODEL D 5F AR MODIFIED (5PK) 20155214;  MODEL D 5F AR MODIFIED (5PK) 20155215;  MODEL D 5F AR MODIFIED (5PK) 20162440;  MODEL D 5F AR MODIFIED (5PK) 20162441; MODEL D 5F AR MODIFIED (5PK) 20162442;  MODEL D 5F AR MODIFIED (5PK) 20162443;  MODEL D 5F AR MODIFIED (5PK) 20162444;  MODEL D 5F AR MODIFIED (5PK) 20167780;  MODEL D 5F AR MODIFIED (5PK) 20167781;  MODEL D 5F AR MODIFIED (5PK) 20167782;  MODEL D 5F AR MODIFIED (5PK) 20174341;  MODEL D 5F AR MODIFIED (5PK) 20178127;  MODEL D 5F AR MODIFIED (5PK) 20178128;  MODEL D 5F AL1 SINGLE 20154070;  MODEL D 5F AL1 SINGLE 20158221;  MODEL D 5F AL1 SINGLE 20162601;  MODEL D 5F AL1 SINGLE 20168996;  MODEL D 5F AL1 SINGLE 20174509;  MODEL D 5F AL1 (5PK) 20153952;  MODEL D 5F AL1 (5PK) 20155216;  MODEL D 5F AL1 (5PK) 20160760;  MODEL D 5F AL1 (5PK) 20162445;  MODEL D 5F AL1 (5PK) 20165784;  MODEL D 5F AL1 (5PK) 20167783;  MODEL D 5F AL1 (5PK) 20170565;  MODEL D 5F AL1 (5PK) 20172903;  MODEL D 5F AL1 (5PK) 20174342;  MODEL D 5F AL1 (5PK) 20178129;  MODEL D 5F AL1 (5PK) 20181279;  MODEL D 5F AL2 (5PK) 20155217;  MODEL D 5F AL2 (5PK) 20162446;  MODEL D 5F AL2 (5PK) 20167784;  MODEL D 5F AL2 (5PK) 20168869;  MODEL D 5F AL2 (5PK) 20178130;  MODEL D 5F AL2 (5PK) 20179497;  MODEL D 5F AL2 (5PK) 20181280;  MODEL D 6F FR3.5 SINGLE 20156910;  MODEL D 6F FR3.5 SINGLE 20167940;  MODEL D 6F FR3.5 SINGLE 20170699;  MODEL D 6F FR3.5 (5PK) 20155218;  MODEL D 6F FR3.5 (5PK) 20158103;  MODEL D 6F FR3.5 (5PK) 20162447;  MODEL D 6F FR3.5 (5PK) 20164319;  MODEL D 6F FR3.5 (5PK) 20165785;  MODEL D 6F FR3.5 (5PK) 20172904;  MODEL D 6F FR3.5 (5PK) 20174343;  MODEL D 6F FR3.5 (5PK) 20178131;  MODEL D 6F FR3.5 (5PK) 20179498;  MODEL D 6F FR3.5 (5PK) 20179499;  MODEL D 6F FR4 SINGLE 20155425;  MODEL D 6F FR4 SINGLE 20156911;  MODEL D 6F FR4 SINGLE 20160869;  MODEL D 6F FR4 SINGLE 20162602;  MODEL D 6F FR4 SINGLE 20164496;  MODEL D 6F FR4 SINGLE 20167941;  MODEL D 6F FR4 SINGLE 20170700;  MODEL D 6F FR4 SINGLE 20173020;  MODEL D 6F FR4 SINGLE 20174510;  MODEL D 6F FR4 SINGLE 20178284;  MODEL D 6F FR4 SINGLE 20179660;  MODEL D 6F FR4 (5PK) 12761862;  MODEL D 6F FR4 (5PK) 20155220;  MODEL D 6F FR4 (5PK) 20155221;  MODEL D 6F FR4 (5PK) 20155222;  MODEL D 6F FR4 (5PK) 20155223;  MODEL D 6F FR4 (5PK) 20155224;  MODEL D 6F FR4 (5PK) 20155225;  MODEL D 6F FR4 (5PK) 20155226;  MODEL D 6F FR4 (5PK) 20155227;  MODEL D 6F FR4 (5PK) 20156794;  MODEL D 6F FR4 (5PK) 20156795;  MODEL D 6F FR4 (5PK) 20156796;  MODEL D 6F FR4 (5PK) 20156798;  MODEL D 6F FR4 (5PK) 20156799;  MODEL D 6F FR4 (5PK) 20156800;  MODEL D 6F FR4 (5PK) 20156801;  MODEL D 6F FR4 (5PK) 20156802;  MODEL D 6F FR4 (5PK) 20156803;  MODEL D 6F FR4 (5PK) 20158104;  MODEL D 6F FR4 (5PK) 20162448;  MODEL D 6F FR4 (5PK) 20162449;  MODEL D 6F FR4 (5PK) 20162450;  MODEL D 6F FR4 (5PK) 20162451;  MODEL D 6F FR4 (5PK) 20162452;  MODEL D 6F FR4 (5PK) 20162453;  MODEL D 6F FR4 (5PK) 20164320;  MODEL D 6F FR4 (5PK) 20164321;  MODEL D 6F FR4 (5PK) 20164322;  MODEL D 6F FR4 (5PK) 20164323;  MODEL D 6F FR4 (5PK) 20164324;  MODEL D 6F FR4 (5PK) 20164325;  MODEL D 6F FR4 (5PK) 20164326;  MODEL D 6F FR4 (5PK) 20164327;  MODEL D 6F FR4 (5PK) 20164328;  MODEL D 6F FR4 (5PK) 20165786;  MODEL D 6F FR4 (5PK) 20165787;  MODEL D 6F FR4 (5PK) 20165788;  MODEL D 6F FR4 (5PK) 20165789;  MODEL D 6F FR4 (5PK) 20165790;  MODEL D 6F FR4 (5PK) 20165791;  MODEL D 6F FR4 (5PK) 20165792;  MODEL D 6F FR4 (5PK) 20165793;  MODEL D 6F FR4 (5PK) 20165794;  MODEL D 6F FR4 (5PK) 20167785;  MODEL D 6F FR4 (5PK) 20167786;  MODEL D 6F FR4 (5PK) 20167787;  MODEL D 6F FR4 (5PK) 20167788;  MODEL D 6F FR4 (5PK) 20167789;  MODEL D 6F FR4 (5PK) 20167790;  MODEL D 6F FR4 (5PK) 20167791;  MODEL D 6F FR4 (5PK) 20167792;  MODEL D 6F FR4 (5PK) 20168870;  MODEL D 6F FR4 (5PK) 20168872;  MODEL D 6F FR4 (5PK) 20168873;  MODEL D 6F FR4 (5PK) 20170566;  MODEL D 6F FR4 (5PK) 20170567;  MODEL D 6F FR4 (5PK) 20170568;  MODEL D 6F FR4 (5PK) 20170569;  MODEL D 6F FR4 (5PK) 20170570;  MODEL D 6F FR4 (5PK) 20170571;  MODEL D 6F FR4 (5PK) 20170572;  MODEL D 6F FR4 (5PK) 20170573;  MODEL D 6F FR4 (5PK) 20172905;  MODEL D 6F FR4 (5PK) 20172906;  MODEL D 6F FR4 (5PK) 20172907;  MODEL D 6F FR4 (5PK) 20172910;  MODEL D 6F FR4 (5PK) 20172911;  MODEL D 6F FR4 (5PK) 20172912;  MODEL D 6F FR4 (5PK) 20174344;  MODEL D 6F FR4 (5PK) 20174345;  MODEL D 6F FR4 (5PK) 20174346;  MODEL D 6F FR4 (5PK) 20174349;  MODEL D 6F FR4 (5PK) 20174350;  MODEL D 6F FR4 (5PK) 20174351;  MODEL D 6F FR4 (5PK) 20174352;  MODEL D 6F FR4 (5PK) 20178132;  MODEL D 6F FR4 (5PK) 20178136;  MODEL D 6F FR4 (5PK) 20178137;  MODEL D 6F FR4 (5PK) 20179500;  MODEL D 6F FR4 (5PK) 20179501;  MODEL D 6F FR4 (5PK) 20179502;  MODEL D 6F FR4 (5PK) 20179504;  MODEL D 6F FR4 (5PK) 20181281;  MODEL D 6F FR4 (5PK) 20181282;  MODEL D 6F FR4 (5PK) 20181283;  MODEL D 6F FR4 (5PK) 20181284;  MODEL D 6F FR4 (5PK) 20181285;  MODEL D 6F FR4 (5PK) 20181286;  MODEL D 6F FR4 (5PK) 20181287;  MODEL D 6F FR4 (5PK) 20183155;  MODEL D 6F FR4 (5PK) 20183156;  MODEL D 6F FR4 (5PK) 20183157;  MODEL D 6F FR4 (5PK) 20183158;  MODEL - 6F EXPO MOD FR4 5-PK 20160644;  MODEL D 6F FR5 (5PK) 20155228;  MODEL D 6F FR5 (5PK) 20156804;  MODEL D 6F FR5 (5PK) 20160763;  MODEL D 6F FR5 (5PK) 20164329;  MODEL D 6F FR5 (5PK) 20165795;  MODEL D 6F FR5 (5PK) 20167793;  MODEL D 6F FR5 (5PK) 20178138;  MODEL D 6F FR6 (5PK) 20162454;  MODEL D 6F FR4 MODIFIED (5PK) 20160764;  MODEL D 6F AL3 SINGLE 20174511;  MODEL D 6F AL3 (5PK) 20167794;  MODEL D 6F AL3 (5PK) 20179506;  MODEL D 6F MPA-1 SINGLE 20162603;  MODEL D 6F MPA-1 SINGLE 20164497;  MODEL D 6F MPA-1 SINGLE 20174512;  MODEL D 6F MPA-1 W/O SH (5PK) 20155229;  MODEL D 6F MPA-1 W/O SH (5PK) 20156805;  MODEL D 6F MPA-1 W/O SH (5PK) 20160765;  MODEL D 6F MPA-1 W/O SH (5PK) 20162455;  MODEL D 6F MPA-1 W/O SH (5PK) 20164330;  MODEL D 6F MPA-1 W/O SH (5PK) 20165796;  MODEL D 6F MPA-1 W/O SH (5PK) 20167795;  MODEL D 6F MPA-1 W/O SH (5PK) 20170576;  MODEL D 6F MPA-1 W/O SH (5PK) 20174354;  MODEL D 6F MPA-1 W/O SH (5PK) 20178139;  MODEL D 6F MPA-1 W/O SH (5PK) 20179507;  MODEL D 6F MPA-2 SH SINGLE 20164498;  MODEL D 6F MPA-2 W/SH (5PK) 20155230;  MODEL D 6F MPA-2 W/SH (5PK) 20156806;  MODEL D 6F MPA-2 W/SH (5PK) 20162456;  MODEL D 6F MPA-2 W/SH (5PK) 20164331;  MODEL D 6F MPA-2 W/SH (5PK) 20168880;  MODEL D 6F MPA-2 W/SH (5PK) 20170577;  MODEL D 6F MPA-2 W/SH (5PK) 20174355;  MODEL D 6F MPA-2 W/SH (5PK) 20178140;  MODEL D 6F MPB-1 SINGLE 20160870;  MODEL D 6F MPB-1 W/O SH (5PK) 20155231;  MODEL D 6F MPB-2 SH SINGLE 20156912;  MODEL D 6F MPB-2 SH SINGLE 20160871;  MODEL D 6F MPB-2 W/SH (5PK) 20160767;  MODEL D 6F MPB-2 W/SH (5PK) 20164332;  MODEL D 6F SON2 100CM SH (5PK) 20164336;  MODEL D 6F SON2 100CM SH (5PK) 20164338;  MODEL D 6F SON2 100CM SH (5PK) 20164339;  MODEL D 6F SON2 100CM SH (5PK) 20164340;  MODEL D 6F SON1 100CM (5PK) 20174357; MODEL D 6F SON1 100CM (5PK) 20174358; MODEL D 6F SON1 100CM (5PK) 20174359;  MODEL D 6F SON1 100CM (5PK) 20174361;  MODEL D 6F CAS1 (5PK) 20164343;  MODEL D 6F CAS1 (5PK) 20178141;  MODEL D 6F CAS2 (5PK) 20164344;  MODEL - EXPO 6F SELECTIVE KIMNY SINGLE 20160872;  MODEL - EXPO 6F SELECTIVE KIMNY 5 PACK 20156807;  MODEL D 6F RCB SINGLE 20164499; MODEL D 6F RCB SINGLE 20165906;  MODEL D 6F RCB SINGLE 20179661;  MODEL D 6F RCB (5PK) 20155233;  MODEL D 6F RCB (5PK) 20155234;  MODEL D 6F RCB (5PK) 20155235;  MODEL D 6F RCB (5PK) 20155237;  MODEL D 6F RCB (5PK) 20155238;  MODEL D 6F RCB (5PK) 20155239;  MODEL D 6F RCB (5PK) 20158115;  MODEL D 6F RCB (5PK) 20158116;  MODEL D 6F RCB (5PK) 20158117;  MODEL D 6F RCB (5PK) 20158118;  MODEL D 6F RCB (5PK) 20158119;  MODEL D 6F RCB (5PK) 20158120;  MODEL D 6F RCB (5PK) 20158121;  MODEL D 6F RCB (5PK) 20162457;  MODEL D 6F RCB (5PK) 20162458;  MODEL D 6F RCB (5PK) 20162459;  MODEL D 6F RCB (5PK) 20162460;  MODEL D 6F RCB (5PK) 20162461;  MODEL D 6F RCB (5PK) 20162462;  MODEL D 6F RCB (5PK) 20162463;  MODEL D 6F RCB (5PK) 20164345;  MODEL D 6F RCB (5PK) 20164346;  MODEL D 6F RCB (5PK) 20164347;  MODEL D 6F RCB (5PK) 20164348;  MODEL D 6F RCB (5PK) 20164349;  MODEL D 6F RCB (5PK) 20164350;  MODEL D 6F RCB (5PK) 20164351;  MODEL D 6F RCB (5PK) 20167798;  MODEL D 6F RCB (5PK) 20167799;  MODEL D 6F RCB (5PK) 20167800;  MODEL D 6F RCB (5PK) 20167801;  MODEL D 6F RCB (5PK) 20167802;  MODEL D 6F RCB (5PK) 20167803;  MODEL D 6F RCB (5PK) 20167804;  MODEL D 6F RCB (5PK) 20170578;  MODEL D 6F RCB (5PK) 20170580;  MODEL D 6F RCB (5PK) 20170581;  MODEL D 6F RCB (5PK) 20170584;  MODEL D 6F RCB (5PK) 20174366;  MODEL D 6F RCB (5PK) 20174367;  MODEL D 6F RCB (5PK) 20174368;  MODEL D 6F RCB (5PK) 20174369;  MODEL D 6F RCB (5PK) 20174370;  MODEL D 6F RCB (5PK) 20174371;  MODEL D 6F RCB (5PK) 20174372;  MODEL D 6F RCB (5PK) 20178142;  MODEL D 6F RCB (5PK) 20178143;  MODEL D 6F RCB (5PK) 20178144;  MODEL D 6F RCB (5PK) 20178146;  MODEL D 6F RCB (5PK) 20179509;  MODEL D 6F RCB (5PK) 20179510;  MODEL D 6F RCB (5PK) 20179511;  MODEL D 6F RCB (5PK) 20179512;  MODEL D 6F RCB (5PK) 20179515;  MODEL D 6F LCB SINGLE 20158224;  MODEL D 6F LCB SINGLE 20164500;  MODEL D 6F LCB SINGLE 20167942;  MODEL D 6F LCB SINGLE 20168998;  MODEL D 6F LCB (5PK) 20155240;  MODEL D 6F LCB (5PK) 20156808;  MODEL D 6F LCB (5PK) 20162464;  MODEL D 6F LCB (5PK) 20164352;  MODEL D 6F LCB (5PK) 20167805;  MODEL D 6F LCB (5PK) 20170585;  MODEL D 6F LCB (5PK) 20174373;  MODEL D 6F LCB (5PK) 20178149;  MODEL D 6F IM SINGLE 20155426;  MODEL D 6F IM SINGLE 20156914;  MODEL D 6F IM SINGLE 20164501;  MODEL D 6F IM (5PK) 20155241; MODEL D 6F IM (5PK) 20156809; MODEL D 6F IM (5PK) 20158122; MODEL D 6F IM (5PK) 20162465;  MODEL D 6F IM (5PK) 20162466;  MODEL D 6F IM (5PK) 20162467;  MODEL D 6F IM (5PK) 20164353;  MODEL D 6F IM (5PK) 20165798;  MODEL D 6F IM (5PK) 20167806;  MODEL D 6F IM (5PK) 20170586;  MODEL D 6F IM (5PK) 20172915;  MODEL D 6F IM (5PK) 20174374;  MODEL D 6F IM (5PK) 20178150;  MODEL D 6F IM (5PK) 20179516;  MODEL D 6F IM (5PK) 20181290;  MODEL - D 6F IM T (SGL) 20155427; MODEL - D 6F I/M T (5 PK) 20155242;  MODEL - D 6F I/M T (5 PK) 20155243;  MODEL - D 6F I/M T (5 PK) 20155244;  MODEL - D 6F I/M T (5 PK) 20155245;  MODEL - D 6F I/M T (5 PK) 20155246;  MODEL - D 6F I/M T (5 PK) 20155247;  MODEL - D 6F I/M T (5 PK) 20155248;  MODEL - D 6F I/M T (5 PK) 20162468;  MODEL - D 6F I/M T (5 PK) 20162470;  MODEL - D 6F I/M T (5 PK) 20162471;  MODEL - D 6F I/M T (5 PK) 20164354;  MODEL - D 6F I/M T (5 PK) 20164355;  MODEL - D 6F I/M T (5 PK) 20164356;  MODEL - D 6F I/M T (5 PK) 20164357;  MODEL - D 6F I/M T (5 PK) 20170588;  MODEL - D 6F I/M T (5 PK) 20170589;  MODEL - D 6F I/M T (5 PK) 20170590;  MODEL - D 6F I/M T (5 PK) 20174375;  MODEL - D 6F I/M T (5 PK) 20174376;  MODEL - D 6F I/M T (5 PK) 20174377;  MODEL - D 6F I/M T (5 PK) 20174378;  MODEL-6F EXPO IMC SINGLE 20158225;  MODEL-6F EXPO IMC SINGLE 20164502;  MODEL-6F EXPO IMC SINGLE 20167944;  MODEL-6F EXPO IMC 5 PACK 20158125;  MODEL-6F EXPO IMC 5 PACK 20167807;  MODEL-6F EXPO IMC 5 PACK 20167808;  MODEL-6F EXPO IMC 5 PACK 20178151;  MODEL-6F EXPO IMC 5 PACK 20178152;  MODEL - 6F EXPO FL3 5-PK 20160641;  MODEL D 6F FL3.5 SINGLE 20160873;  MODEL D 6F FL3.5 SINGLE 20164503;  MODEL D 6F FL3.5 SINGLE 20167945;  MODEL D 6F FL3.5 (5PK) 20155250;  MODEL D 6F FL3.5 (5PK) 20155251;  MODEL D 6F FL3.5 (5PK) 20156810;  MODEL D 6F FL3.5 (5PK) 20156811;  MODEL D 6F FL3.5 (5PK) 20158126;  MODEL D 6F FL3.5 (5PK) 20162472;  MODEL D 6F FL3.5 (5PK) 20162473;  MODEL D 6F FL3.5 (5PK) 20164358;  MODEL D 6F FL3.5 (5PK) 20164359;  MODEL D 6F FL3.5 (5PK) 20165799;  MODEL D 6F FL3.5 (5PK) 20165800;  MODEL D 6F FL3.5 (5PK) 20167809;  MODEL D 6F FL3.5 (5PK) 20168884;  MODEL D 6F FL3.5 (5PK) 20170595;  MODEL D 6F FL3.5 (5PK) 20172916;  MODEL D 6F FL3.5 (5PK) 20174382;  MODEL D 6F FL3.5 (5PK) 20178153;  MODEL D 6F FL3.5 (5PK) 20178154;  MODEL D 6F FL3.5 (5PK) 20179517;  MODEL D 6F FL3.5 (5PK) 20181292;  MODEL D 6F FL4 SINGLE 20155428;  MODEL D 6F FL4 SINGLE 20156915;  MODEL D 6F FL4 SINGLE 20158228;  MODEL D 6F FL4 SINGLE 20164505;  MODEL D 6F FL4 SINGLE 20167946;  MODEL D 6F FL4 SINGLE 20168999;  MODEL D 6F FL4 SINGLE 20174517;  MODEL D 6F FL4 SINGLE 20178286;  MODEL D 6F FL4 SINGLE 20179663;  MODEL D 6F FL4 (5PK) 20155255;  MODEL D 6F FL4 (5PK) 20156812;  MODEL D 6F FL4 (5PK) 20156813;  MODEL D 6F FL4 (5PK) 20156814;  MODEL D 6F FL4 (5PK) 20156815;  MODEL D 6F FL4 (5PK) 20156816;  MODEL D 6F FL4 (5PK) 20156817;  MODEL D 6F FL4 (5PK) 20156818;  MODEL D 6F FL4 (5PK) 20158128;  MODEL D 6F FL4 (5PK) 20158129;  MODEL D 6F FL4 (5PK) 20160773;  MODEL D 6F FL4 (5PK) 20160774;  MODEL D 6F FL4 (5PK) 20160775;  MODEL D 6F FL4 (5PK) 20160776;  MODEL D 6F FL4 (5PK) 20160777;  MODEL D 6F FL4 (5PK) 20160778;  MODEL D 6F FL4 (5PK) 20160779;  MODEL D 6F FL4 (5PK) 20160780;  MODEL D 6F FL4 (5PK) 20160781;  MODEL D 6F FL4 (5PK) 20160782;  MODEL D 6F FL4 (5PK) 20162475;  MODEL D 6F FL4 (5PK) 20162476;  MODEL D 6F FL4 (5PK) 20162477;  MODEL D 6F FL4 (5PK) 20162478;  MODEL D 6F FL4 (5PK) 20162479;  MODEL D 6F FL4 (5PK) 20162480;  MODEL D 6F FL4 (5PK) 20162481;  MODEL D 6F FL4 (5PK) 20162482;  MODEL D 6F FL4 (5PK) 20162483;  MODEL D 6F FL4 (5PK) 20162484;  MODEL D 6F FL4 (5PK) 20162485;  MODEL D 6F FL4 (5PK) 20162486;  MODEL D 6F FL4 (5PK) 20162487;  MODEL D 6F FL4 (5PK) 20164360;  MODEL D 6F FL4 (5PK) 20164361;  MODEL D 6F FL4 (5PK) 20164362;  MODEL D 6F FL4 (5PK) 20164363;  MODEL D 6F FL4 (5PK) 20164364;  MODEL D 6F FL4 (5PK) 20164365;  MODEL D 6F FL4 (5PK) 20164366;  MODEL D 6F FL4 (5PK) 20165801;  MODEL D 6F FL4 (5PK) 20165802;  MODEL D 6F FL4 (5PK) 20165803;  MODEL D 6F FL4 (5PK) 20165804;  MODEL D 6F FL4 (5PK) 20167810;  MODEL D 6F FL4 (5PK) 20167811;  MODEL D 6F FL4 (5PK) 20167812;  MODEL D 6F FL4 (5PK) 20167813;  MODEL D 6F FL4 (5PK) 20167814;  MODEL D 6F FL4 (5PK) 20167815;  MODEL D 6F FL4 (5PK) 20167816;  MODEL D 6F FL4 (5PK) 20170596;  MODEL D 6F FL4 (5PK) 20170597;  MODEL D 6F FL4 (5PK) 20170598;  MODEL D 6F FL4 (5PK) 20170599;  MODEL D 6F FL4 (5PK) 20170600;  MODEL D 6F FL4 (5PK) 20170601;  MODEL D 6F FL4 (5PK) 20170602;  MODEL D 6F FL4 (5PK) 20170603;  MODEL D 6F FL4 (5PK) 20172917;  MODEL D 6F FL4 (5PK) 20172918;  MODEL D 6F FL4 (5PK) 20172919;  MODEL D 6F FL4 (5PK) 20172920;  MODEL D 6F FL4 (5PK) 20172921;  MODEL D 6F FL4 (5PK) 20172922;  MODEL D 6F FL4 (5PK) 20172923;  MODEL D 6F FL4 (5PK) 20172924;  MODEL D 6F FL4 (5PK) 20172925;  MODEL D 6F FL4 (5PK) 20172926;  MODEL D 6F FL4 (5PK) 20174383;  MODEL D 6F FL4 (5PK) 20174384;  MODEL D 6F FL4 (5PK) 20174385;  MODEL D 6F FL4 (5PK) 20174386;  MODEL D 6F FL4 (5PK) 20174387;  MODEL D 6F FL4 (5PK) 20174388;  MODEL D 6F FL4 (5PK) 20174389;  MODEL D 6F FL4 (5PK) 20174390;  MODEL D 6F FL4 (5PK) 20178155;  MODEL D 6F FL4 (5PK) 20178156;  MODEL D 6F FL4 (5PK) 20178157;  MODEL D 6F FL4 (5PK) 20178158;  MODEL D 6F FL4 (5PK) 20178159;  MODEL D 6F FL4 (5PK) 20178160;  MODEL D 6F FL4 (5PK) 20178161;  MODEL D 6F FL4 (5PK) 20178162;  MODEL D 6F FL4 (5PK) 20179518;  MODEL D 6F FL4 (5PK) 20179519;  MODEL D 6F FL4 (5PK) 20179520;  MODEL D 6F FL4 (5PK) 20179521;  MODEL D 6F FL4 (5PK) 20179522;  MODEL D 6F FL4 (5PK) 20181294;  MODEL D 6F FL4 (5PK) 20181295;  MODEL D 6F FL4 (5PK) 20181296;  MODEL D 6F FL4 (5PK) 20181299;  MODEL D 6F FL4.5 SINGLE 20178287;  MODEL D 6F FL4.5 (5PK) 20156819;  MODEL D 6F FL4.5 (5PK) 20160785;  MODEL D 6F FL5 SINGLE 20155429;  MODEL D 6F FL5 SINGLE 20164506;  MODEL D 6F FL5 SINGLE 20171718;  MODEL D 6F FL5 (5PK) 20155259;  MODEL D 6F FL5 (5PK) 20156820;  MODEL D 6F FL5 (5PK) 20158137;  MODEL D 6F FL5 (5PK) 20162489;  MODEL D 6F FL5 (5PK) 20164368;  MODEL D 6F FL5 (5PK) 20165805;  MODEL D 6F FL5 (5PK) 20167817;  MODEL D 6F FL5 (5PK) 20174391;  MODEL D 6F FL5 (5PK) 20178163;  MODEL D 6F FL5 (5PK) 20179523;  MODEL D 6F FL5 (5PK) 20181303;  MODEL D 6F FL6 SINGLE 20164507;  MODEL D 6F FL6 (5PK) 20155260;  MODEL D 6F FL6 (5PK) 20156821;  MODEL D 6F FL6 (5PK) 20162490;  MODEL D 6F FL6 (5PK) 20164369;  MODEL D 6F FL6 (5PK) 20167818;  MODEL D 6F FL6 (5PK) 20179524;  MODEL D 6F (5PKX3) 20181304 08641-300; MODEL D 6F PIG 6 HOLE SINGLE 20162605; MODEL D 6F PIG 6 HOLE SINGLE 20178289; MODEL D 6F PIG 6 HOLE (5PK) 20156829; MODEL D 6F PIG 6 HOLE (5PK) 20158144;  MODEL D 6F PIG 6 HOLE (5PK) 20158145;  MODEL D 6F PIG 6 HOLE (5PK) 20158146;  MODEL D 6F PIG 6 HOLE (5PK) 20158147;  MODEL D 6F PIG 6 HOLE (5PK) 20158148;  MODEL D 6F PIG 6 HOLE (5PK) 20160791;  MODEL D 6F PIG 6 HOLE (5PK) 20160792;  MODEL D 6F PIG 6 HOLE (5PK) 20160793;  MODEL D 6F PIG 6 HOLE (5PK) 20160794;  MODEL D 6F PIG 6 HOLE (5PK) 20160795;  MODEL D 6F PIG 6 HOLE (5PK) 20162496;  MODEL D 6F PIG 6 HOLE (5PK) 20162497;  MODEL D 6F PIG 6 HOLE (5PK) 20162498;  MODEL D 6F PIG 6 HOLE (5PK) 20162499;  MODEL D 6F PIG 6 HOLE (5PK) 20162500;  MODEL D 6F PIG 6 HOLE (5PK) 20164375;  MODEL D 6F PIG 6 HOLE (5PK) 20165810;  MODEL D 6F PIG 6 HOLE (5PK) 20168900;  MODEL D 6F PIG 6 HOLE (5PK) 20168901;  MODEL D 6F PIG 6 HOLE (5PK) 20168902;  MODEL D 6F PIG 6 HOLE (5PK) 20170612;  MODEL D 6F PIG 6 HOLE (5PK) 20172934;  MODEL D 6F PIG 6 HOLE (5PK) 20172935;  MODEL D 6F PIG 6 HOLE (5PK) 20172936;  MODEL D 6F PIG 6 HOLE (5PK) 20172937;  MODEL D 6F PIG 6 HOLE (5PK) 20174399;  MODEL D 6F PIG 6 HOLE (5PK) 20174400;  MODEL D 6F PIG 6 HOLE (5PK) 20174401;  MODEL D 6F PIG 6 HOLE (5PK) 20174402;  MODEL D 6F PIG 6 HOLE (5PK) 20178169;  MODEL D 6F PIG 6 HOLE (5PK) 20178170;  MODEL D 6F PIG 6 HOLE (5PK) 20178171;  MODEL D 6F PIG 6 HOLE (5PK) 20178172;  MODEL D 6F PIG 6 HOLE (5PK) 20179533;  MODEL D 6F PIG 6 HOLE (5PK) 20179534;  MODEL D 6F PIG 6 HOLE (5PK) 20179535;  MODEL D 6F 145 PIG 110 SINGLE 20178290;  MODEL D 6F 145 PIG 110CM (5PK) 20156832;  MODEL D 6F 145 PIG 110CM (5PK) 20156833;  MODEL D 6F 145 PIG 110CM (5PK) 20158149;  MODEL D 6F 145 PIG 110CM (5PK) 20158150;  MODEL D 6F 145 PIG 110CM (5PK) 20158151;  MODEL D 6F 145 PIG 110CM (5PK) 20160796;  MODEL D 6F 145 PIG 110CM (5PK) 20168903;  MODEL D 6F 145 PIG 110CM (5PK) 20168904;  MODEL D 6F 145 PIG 110CM (5PK) 20170615;  MODEL D 6F 145 PIG 110CM (5PK) 20170616;  MODEL D 6F 145 PIG 110CM (5PK) 20172938;  MODEL D 6F 145 PIG 110CM (5PK) 20172939;  MODEL D 6F 145 PIG 110CM (5PK) 20174403;  MODEL D 6F 145 PIG 110CM (5PK) 20174404;  MODEL D 6F 145 PIG 110CM (5PK) 20178176;  MODEL D 6F 145 PIG 110CM (5PK) 20179537;  MODEL D 6F 145 PIG 110CM (5PK) 20181313;  MODEL D 6F 145 PIG 110CM (5PK) 20181314;  MODEL D 6F 145 PIG 110CM (5PK) 20183190;  MODEL D 6F 155 PIG 110 CM (5PK) 20156834;  MODEL D 6F 155 PIG 110 CM (5PK) 20156835;  MODEL D 6F 155 PIG 110 CM (5PK) 20158152;  MODEL D 6F 155 PIG 110 CM (5PK) 20167830;  MODEL D 6F 155 PIG 110 CM (5PK) 20170617;  MODEL D 6F 155 PIG 110 CM (5PK) 20172940;  MODEL D 6F 155 PIG 110 CM (5PK) 20174405;  MODEL D 6F 155 PIG 110 CM (5PK) 20178177;  MODEL D 6F 155 PIG 110 CM (5PK) 20181316;  MODEL D 6F AR1 (5PK) 20156836; MODEL D 6F AR1 (5PK) 20162501;  MODEL D 6F AR1 (5PK) 20164381;  MODEL D 6F AR1 (5PK) 20167831;  MODEL D 6F AR1 (5PK) 20168905;  MODEL D 6F AR1 (5PK) 20174407;  MODEL D 6F AR1 (5PK) 20178178;  MODEL D 6F AR1 (5PK) 20179539;  MODEL D 6F AR2 SINGLE 20164509;  MODEL D 6F AR2 SINGLE 20167949;  MODEL D 6F AR2 (5PK) 20155277;  MODEL D 6F AR2 (5PK) 20156837;  MODEL D 6F AR2 (5PK) 20164382;  MODEL D 6F AR2 (5PK) 20170620;  MODEL D 6F AR2 (5PK) 20178179;  MODEL D 6F AR2 (5PK) 20179540;  MODEL D 6F AR MOD SINGLE 20156917;  MODEL D 6F AR MOD SINGLE 20158231;  MODEL D 6F AR MOD SINGLE 20165907;  MODEL D 6F AR MOD SINGLE 20169001;  MODEL D 6F AR MODIFIED (5PK) 20155278;  MODEL D 6F AR MODIFIED (5PK) 20156838;  MODEL D 6F AR MODIFIED (5PK) 20160798;  MODEL D 6F AR MODIFIED (5PK) 20162502;  MODEL D 6F AR MODIFIED (5PK) 20164383;  MODEL D 6F AR MODIFIED (5PK) 20167832;  MODEL D 6F AR MODIFIED (5PK) 20174409;  MODEL D 6F AR MODIFIED (5PK) 20179541;  MODEL D 6F AL1 SINGLE 20154485;  MODEL D 6F AL1 SINGLE 20162606;  MODEL D 6F AL1 SINGLE 20167950;  MODEL D 6F AL1 (5PK) 20155279;  MODEL D 6F AL1 (5PK) 20156839;  MODEL D 6F AL1 (5PK) 20160799;  MODEL D 6F AL1 (5PK) 20162503;  MODEL D 6F AL1 (5PK) 20164384;  MODEL D 6F AL1 (5PK) 20165815;  MODEL D 6F AL1 (5PK) 20167833;  MODEL D 6F AL1 (5PK) 20168907;  MODEL D 6F AL1 (5PK) 20170621;  MODEL D 6F AL1 (5PK) 20174410;  MODEL D 6F AL1 (5PK) 20178181;  MODEL D 6F AL1 (5PK) 20179542;  MODEL D 6F AL2 SINGLE 20173023;  MODEL D 6F AL2 (5PK) 20155280;  MODEL D 6F AL2 (5PK) 20156840;  MODEL D 6F AL2 (5PK) 20160800;  MODEL D 6F AL2 (5PK) 20162504;  MODEL D 6F AL2 (5PK) 20164385;  MODEL D 6F AL2 (5PK) 20167834;  MODEL D 6F AL2 (5PK) 20174411;  MODEL D 6F AL2 (5PK) 20178182;  MODEL D 6F AL2 (5PK) 20179543;  MODEL-6F EXPO AL2 65cm 5-PK 20163164;  MODEL D 7F MPA-1 W/O SH (5PK) 20144187;  MODEL D 7F MPB-2 W/ SH (5PK) 20158156;  MODEL D 7F MPB-2 W/ SH (5PK) 20164386;  MODEL D 7F IM (5PK) 20155283;  MODEL D 7F FL4 (5PK) 20179545;  MODEL D 7F PIG 110 CM (5PK) 20158157;  MODEL D 7F PIG 110 CM (5PK) 20165818;  MODEL D 7F PIG 110 CM (5PK) 20167836;  MODEL D 7F PIG 145 110 CM (5PK) 20165819;  MODEL D 7F PIG 145 110 CM (5PK) 20167837;  MODEL D 7F PIG 145 110 CM (5PK) 20174416;  MODEL D 7F PIG 155 110 CM (5PK) 20162505; and  MODEL D 7F PIG 155 110 CM (5PK) 20167838.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- US (including DC and Puerto Rico), Japan, United Arab Emirates, Albania, Austria, Belgium, Bulgaria, Bahrain, Switzerland, Czech Republic, Germany, Denmark, Algeria, Egypt, Spain, Finland, France, Great Britain, Greece, Hungary, Ireland, Israel, Iraq, Iceland, Italy, Jordan, Kuwait, Lebanon, Lithuania, Libya, Morocco, Netherlands, Norway, Pakistan, Poland, West Bank Gaza Strip/ Palestine, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Slovakia, Syria, Turkey, Ukraine, South Africa, Argentina, Australia, Brazil, Canada, Chile, Colombia, Dominican Republic, Ecuador, Guatemala, Indonesia, India, South Korea, Mexico, Malaysia, Nicaragua, New Zealand, Panama, Philippines, Puerto Rico, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, American Virgin Islands and Vietnam.
  • Product Description
    Boston Scientific Expo, Angiographic Catheter, Sterilized with ethylene oxide gas, Made in Mexico, Parque Industrial La Mesa, Edificion G1, Suites B. H, Tijuana, Baja California 22650 MEX. || Boston Scientific Guide Catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in cerebral vasculature.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Boston Scientific Corporation, 1 Scimed Pl, Maple Grove MN 55311-1565
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    USFDA