Recall of Ethicon

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ethicon, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62056
  • Event Risk Class
    Class 2
  • Event Number
    Z-1897-2012
  • Event Initiated Date
    2012-04-03
  • Event Date Posted
    2012-06-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-04-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Agent, absorbable hemostatic, collagen based - Product Code LMF
  • Reason
    The lyophilized thrombin component of the surgiflo hemostatic matrix kit with thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process.
  • Action
    Ethicon Inc. sent a Urgent Voluntary Product Recall letter dated April 3, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you have any product subject to this voluntary recall, please remove it from your inventory immediately and return it per the following instructions. Return product to Stericycle using the enclosed pre-paid shipping label. Shipping fees wil be charged directly to Ethicon through the use of this shipping label. No reimbursements will be made for the shipping charge if this label is not used. Extra shipping labels may be obtained by calling 1-866-338-2611. Credit will be issued for impacted product that is returned. Further questions please call 1-866-338-2611, or contact your Ethicon Biosurgery Sales Representative. Thank you for your coorporation and immediate assistance. .

Device

  • Model / Serial
    D100622   Product Code Lot Number Expiration Date  199102S 237115 8/31/2012 199102S 238568 9/30/2012 199102S 238569 11/30/2012 199102S 237493 6/30/2012 2993 235489 11/30/2012 2993 238599 3/31/2013 2993SAMP 238369 3/31/2013 SAMP199102S 237493 6/30/2012 SAMP199102S 283569 11/30/2012
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    USA (nationwide)
  • Product Description
    Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip || Distributed by Ethicon Ethicon Inc., Somerville, NJ 08876 USA || Manufactured by Ferrosan Medical Devices || Ferrosan Medical Devices A/S || Sydmarken 5, DK-2860 Soeberg, Denmark || Mixed with sterile saline or thrombin solution is indicated for surgical procedures (except opthalmic) for hemostasis when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ethicon, Inc., US Highway 22 West, Somerville NJ 08876
  • Manufacturer Parent Company (2017)
  • Source
    USFDA