Recall of DR Systems Unity PACS software, now known as Merge Unity PACS software.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Merge Healthcare, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    75876
  • Event Risk Class
    Class 2
  • Event Number
    Z-0939-2017
  • Event Initiated Date
    2015-10-15
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, image processing, radiological - Product Code LLZ
  • Reason
    The software fails to associate to the correct mg image if there are two images for the same view.
  • Action
    Merge sent an Urgent : Medical Device Recall letter dated July 11, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Be aware of the following workaround: Customer I User: Ensure both" tome" images are reviewed when more than one is captured for the same view. Merge has a released fix available for this issue. YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED Please reply using the enclosed form and the return addressed envelope. Your response is required no later than August 2,2016. For further questions regarding this recall please call (877) 741-5369

Device

  • Model / Serial
    Version 11.0
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distribution was made to medical facilities located in MT, CA, PA, and TX.
  • Product Description
    DR Systems Unity PACS software, now known as Merge Unity PACS software.
  • Manufacturer