Recall of Dimension Vista HbA1c

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Healthcare Diagnostics, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69334
  • Event Risk Class
    Class 2
  • Event Number
    Z-0127-2015
  • Event Initiated Date
    2014-08-14
  • Event Date Posted
    2014-10-24
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-08-10
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Assay, glycosylated hemoglobin - Product Code LCP
  • Reason
    In march, siemens healthcare diagnostics issued an urgent medical device correction for decreased recovery of patient samples with the dimension vista® csa & csae flex® reagent cartridges when whole blood samples are run from small sample containers (ssc). siemens has confirmed that when whole blood samples are run from ssc, hba1c values may exhibit a percent bias ranging from - 8.2% t.
  • Action
    An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista¿ HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.