Recall of Device Recall Zimmer Unicompartmental Knee System headed screw

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    48716
  • Event Risk Class
    Class 3
  • Event Number
    Z-2300-2008
  • Event Initiated Date
    2008-06-30
  • Event Date Posted
    2008-09-17
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-12-17
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    cutting guide screw - Product Code LXH
  • Reason
    An impurity in the metal may affect the strength of the screw, resulting in breakage and/or surgical delays.
  • Action
    Zimmer instructed consignees by a letter, dated June 30, 2008, sent via mail or email, to return the product.

Device

  • Model / Serial
    Lot 60954069.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Florida, China and India.
  • Product Description
    Zimmer M/G Unicompartmental Knee System headed screw, 48 mm length, sterile, Catalog No. 00-5791-041-00. Used during surgery for the cutting guide. This screw is not permanently implanted.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer Inc., 345 E Main St, Warsaw IN 46580-2746
  • Source
    USFDA