Recall of Device Recall Zimmer

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer Manufacturing B.V..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    72900
  • Event Risk Class
    Class 2
  • Event Number
    Z-1160-2016
  • Event Initiated Date
    2016-01-11
  • Event Date Posted
    2016-03-12
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-04-13
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
  • Reason
    As a result of the insufficient sealer calibration data for product packaged in firm's building ii between august 2010 and april 2013. the affected products are sterile.
  • Action
    Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.

Device

  • Model / Serial
    lot no.: 62313588
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    US Nationwide Distribution
  • Product Description
    Product 62 consists of all product under product code: LPH and same usage: || Item no: || 620003620 TRILOGY ACET SHELL 36MM OD MULTI || For use in total hip arthroplasty
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer Manufacturing B.V., Turpeaux Industrial Park Rd 1 km 123.4, Mercedita PR 00715
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source