Recall of Device Recall Zimmer

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    70500
  • Event Risk Class
    Class 2
  • Event Number
    Z-1219-2015
  • Event Initiated Date
    2015-02-05
  • Event Date Posted
    2015-02-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-09-23
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
  • Reason
    Pmma coating does not meet specifications. area of the stem may be missing pmma coating.
  • Action
    Zimmer sent an URGENT MEDICAL DEVICE RECALL - Lot Specific letter dated February 5, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Your Responsibilities 1. Review the notification and ensure affected personnel are aware of the contents. 2. Locate and remove from inventory the affected products identified above. 3. Return any affected product with missing PMMA coating per the PER process 4. Return the Notification Acknowledgment Form (Attachment 1) to corporatequality.postmarket@zimmer.com. 5. Please notify Zimmer if the hospital that you have distributed the affected product to has implanted the product. In addition, identify the surgeons that have used this product. 6. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.

Device

  • Model / Serial
    Item No. 00-5852-052-10; Lot 62866438
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    US Distribution to the state of TX
  • Product Description
    Zimmer Segmental System (ZSS) Cemented Stem / ZSS Cemented Stem, Smooth Stem || Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non-Porous, Uncemented.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer, Inc., 1800 W Center St, Warsaw IN 46580-2304
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source