Recall of Device Recall VITROS Chemistry Products BuBc Slides

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ortho-Clinical Diagnostics.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65572
  • Event Risk Class
    Class 2
  • Event Number
    Z-0300-2014
  • Event Initiated Date
    2013-06-12
  • Event Date Posted
    2013-11-15
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-04-11
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Diazo colorimetry, bilirubin - Product Code CIG
  • Reason
    Ortho clinical diagnostics is recalling slide dispense cartridges used with some vitros chemistry product slides due to slide dispense issues.
  • Action
    Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.

Device

  • Model / Serial
    Multiple lots affected, with expiry dates through 22-Mar-2015
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.
  • Product Description
    VITROS Chemistry Products BuBc Slides packaged as 300 slides/pack catalog number 8383051 and packaged as 90 slides/pack catalog number 1612365 || VITROS BuBc Slides quantitatively measure unconjugated bilirubin (Bu) and conjugated bilirubin (Bc) concentrations in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ortho-Clinical Diagnostics, 100 Indigo Creek Dr, Rochester NY 14626-5101
  • Manufacturer Parent Company (2017)
  • Source
    USFDA