Recall of Device Recall UNIDOS webline Universal Dosimeter

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by PTW-Freiburg.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    71395
  • Event Risk Class
    Class 2
  • Event Number
    Z-2439-2015
  • Event Initiated Date
    2015-05-27
  • Event Date Posted
    2015-08-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-03-31
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Accelerator, linear, medical - Product Code IYE
  • Reason
    In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. this could result in a radiation over-exposure.
  • Action
    PTW issues Urgent Safety Notice due to an error with firmware version 2.05 of the UNIDOS webline in some cases, especially with the high dose-per-pulse-rates typical for particle therapy, measuring errors can result.

Device

  • Model / Serial
    serial numbers: **T10021: 000005, 000019, 000037, 000041, 000047, 000049, 000058, 000066, 000074, 000076, 000079, 000099, 000107, 000125, 000141, 000144, 000158, 000161, 000168, 000175, 000186, 000212, 000221, 000224, 000229, 000231, 000254, 000262, 000274, 000294, 000301, 000302, 000327, 000351, 000352, 000382, 000401, 000413, 000424, 000437, 000445, 000452, 000470, 000477, 000486, 000504, 000525, 000528, 000535, 000567, 000593, 000594, 000649, 000655, 000658, 000659, 000663, 000678, 000686, 000710, 000711, 000731, 000752, 000756, 000763, 000785, 000792, 000793, 000796, 000800, 000820, 000829, 000830, 000857, 000860, 000863, 000865 ,000871, 000874, 000890, 000892, 000946, **T10022: 000003, 000046, 000066, 000087, 000096, 000106, 000118, 000138, 000172, 000178, 000183, 000186, 000188, 000210, 000240, 000317, 000336, 000347, 000350, 000351, 000358, 000360, 000392, 000398, 000420, 000422, 000425, 000434 -000453, 000456-000469, 000474, ** T10023: 000022, 000028, 000051, 00102, 000196, 000213, 000226, 000250, 000255, 000257, 000259, 000261, 000263, 000271, 000280, 000294
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.
  • Product Description
    PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and health physics.
  • Manufacturer