Recall of Device Recall Ultraview Spacelabs Medical Multigas Analyzer, Model 91518

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Spacelabs Healthcare, Llc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    59136
  • Event Risk Class
    Class 3
  • Event Number
    Z-2902-2011
  • Event Initiated Date
    2011-06-08
  • Event Date Posted
    2011-08-01
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-09-21
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Nitrous-Oxide Gaseous Phase (anesthetic conc.) Analyzer - Product Code CBR
  • Reason
    Spacelabs received a report of the model 91518 multigas analyzer trending low levels of an anesthetic agent that is not present during a particular case. the error is limited to only the trend table data.
  • Action
    Spacelabs Healthcare, Llc. sent a "Customer Bulletin" dated June 20, 2011 to all affected customers. The customers were notified of the trend problem and are advised to follow the instructions in the Perioperative section and the ADMIT/DISCHARGE section of the users manual. For questions contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support or 1-425-657-7200 x5089.

Device

  • Model / Serial
    Model 91518, V1.00.14.  Serial Numbers: 1518-000003, 1518-000008, 1518-000013, 1518-000014, 1518-000016, 1518-000024, 1518-000029, 1518-000036, 1518-000040, 1518-000044, 1518-000226, 1518-000228, 1518-000231, 1518-000233, 1518-000243, 1518-000349, 1518-000354, 1518-000355, 1518-000389, 1518-000399, 1518-000457, 1518-000752, 1518-000819, 1518-000825, 1518-000846, 1518-000948, 1518-000991, 1518-001341, 1518-001393, 1518-001394, 1518-001395, 1518-001522, 1518-001536, 1518-001654, 1518-001655, 1518-001708, 1518-001734, 1518-001736, 1518-001853, 1518-001972, 1518-002065, 1518-002089, 1518-002097, 1518-002114, 1518-002116, 1518-002117, 1518-002118, 1518-002119, 1518-002120, 1518-002121, 1518-002122, 1518-002123, 1518-002124, 1518-002125, 1518-002126, 1518-002127, 1518-002128, 1518-002129, 1518-002130, 1518-002131, 1518-002132, 1518-002133, 1518-002134, 1518-002135, 1518-002136, 1518-002137, 1518-002138, 1518-002139, 1518-002140, 1518-002141, 1518-002142, 1518-002143, 1518-002144, 1518-002145, 1518-002146, 1518-002147, 1518-002148, 1518-002149, 1518-002150, 1518-002151, 1518-002152, 1518-002153, 1518-002154, 1518-002155, 1518-002156, 1518-002157, 1518-002161, 1518-002162, 1518-002163, 1518-002164, 1518-002166, 1518-002167, 1518-002168, 1518-002169, 1518-002170, 1518-002171, 1518-002172, 1518-002173, 1518-002174, 1518-002175, 1518-002176, 1518-002177, 1518-002178, 1518-002179, 1518-002180, 1518-002181, 1518-002182, 1518-002183, 1518-002184, 1518-002185, 1518-002186, 1518-002187, 1518-002188, 1518-002190, 1518-002191, 1518-002192, 1518-002205, 1518-002206, 1518-002207, 1518-002208, 1518-002209, 1518-002210, 1518-002211, 1518-002212, 1518-002213, 1518-002214, 1518-002215, 1518-002216, 1518-002217, 1518-002218, 1518-002219, 1518-002220, 1518-002221, 1518-002222, 1518-002223, 1518-002224, 1518-002225, 1518-002226, 1518-002227, 1518-002228, 1518-002229, 1518-002230, 1518-002231, 1518-002232, 1518-002233, 1518-002234, 1518-002235, 1518-002236, 1518-002237, 1518-002238, 1518-002239, 1518-002240, 1518-002241, 1518-002242, 1518-002243, 1518-002244, 1518-002245, 1518-002246, 1518-002247, 1518-002248, 1518-002249, 1518-002250, 1518-002251, 1518-002252, 1518-002253, 1518-002254, 1518-002255, 1518-002256, 1518-002257, 1518-002258, 1518-002259, 1518-002260, 1518-002261, 1518-002262, 1518-002263, 1518-002264, 1518-002265, 1518-002266, 1518-002267, 1518-002268, 1518-002276, 1518-002277, 1518-002278, 1518-002279, 1518-002280, 1518-002281, 1518-002282, 1518-002283, 1518-002284, 1518-002285, 1518-002286, 1518-002287, 1518-002288, 1518-002289, 1518-002290, 1518-002291, 1518-002292, 1518-002293, 1518-002294, 1518-002295, 1518-002296, 1518-002297, 1518-002298, 1518-002299, 1518-002300, 1518-002301, 1518-002302, 1518-002303, 1518-002304, 1518-002305, 1518-002306, 1518-002307, 1518-002308, 1518-002309, 1518-002310, 1518-002311, 1518-002312, 1518-002313, 1518-002314, 1518-002315, 1518-002316, 1518-002317, 1518-002318, 1518-002319, 1518-002320, 1518-002321, 1518-002322, 1518-002323, 1518-002325, 1518-002326, 1518-002327, 1518-002328, 1518-002329, 1518-002330, 1518-002331, 1518-002332, 1518-002333, 1518-002334, 1518-002335, 1518-002336, 1518-002337, 1518-002338, 1518-002339, 1518-002340, 1518-002341, 1518-002342, 1518-002343, 1518-002344, 1518-002345, 1518-002346, 1518-002347, 1518-002348, 1518-002349, 1518-002350, 1518-002351, 1518-002352, 1518-002353, 1518-002354, 1518-002355, 1518-002356, 1518-002357, 1518-002358, 1518-002359, 1518-002360, 1518-002361, 1518-002362, 1518-002363, 1518-002364, 1518-002365, 1518-002366, 1518-002367, 1518-002368, 1518-002369, 1518-002370, 1518-002371, 1518-002372, 1518-002373, 1518-002374, 1518-002375, 1518-002376, 1518-002377, 1518-002378, 1518-002379, 1518-002380, 1518-002381, 1518-002382, 1518-002383, 1518-002384, 1518-002385, 1518-002386, 1518-002387, 1518-002388, 1518-002389, 1518-002390, 1518-002391, 1518-002392, 1518-002393, 1518-002394, 1518-002395, 1518-002396, 1518-002397, 1518-002398, 1518-002399, 1518-002400, 1518-002401, 1518-002402, 1518-002403, 1518-002404, 1518-002405, 1518-002406, 1518-002407, 1518-002408, 1518-002409, 1518-002410, 1518-002411, 1518-002412, 1518-002413, 1518-002414, 1518-002415, 1518-002416, 1518-002417, 1518-002418, 1518-002419, 1518-002420, 1518-002421, 1518-002422, 1518-002423, 1518-002424, 1518-002425, 1518-002426, 1518-002427, 1518-002428, 1518-002429, 1518-002430, 1518-002431, 1518-002432, 1518-002433, 1518-002434, 1518-002435, 1518-002436, 1518-002437, 1518-002438, 1518-002439, 1518-002440, 1518-002441, 1518-002442, 1518-002443, 1518-002444, 1518-002445, 1518-002446, 1518-002447, 1518-002448, 1518-002451, 1518-002452, 1518-002453, 1518-002454, 1518-002456, 1518-002457, 1518-002458, 1518-002459, 1518-002460, 1518-002461, 1518-002462, 1518-002463, 1518-002464, 1518-002465, 1518-002466, 1518-002467, 1518-002468, 1518-002469, 1518-002470, 1518-002471, 1518-002472, 1518-002476, 1518-002477, 1518-002478, 1518-002479, 1518-002480, 1518-002481, 1518-002482, 1518-002483, 1518-002484, 1518-002485, 1518-002486, 1518-002487, 1518-002488, 1518-002489, 1518-002490, 1518-002491, 1518-002493, 1518-002494, 1518-002495, 1518-002496, 1518-002497, 1518-002498, 1518-002499, 1518-002500, 1518-002501, 1518-002502, 1518-002507, 1518-002508, 1518-002510, 1518-002511, 1518-002513, 1518-002514, 1518-002517, 1518-002518, 1518-002519, 1518-002520, 1518-002521, 1518-002522, 1518-002523, 1518-002524, 1518-002525, 1518-002526, 1518-002527, 1518-002528, 1518-002529, 1518-002530, 1518-002531, 1518-002532, 1518-002533, 1518-002534, 1518-002535, 1518-002536, 1518-002537, 1518-002538, 1518-002539, 1518-002540, 1518-002541, 1518-002542, 1518-002543, 1518-002545, 1518-002549, 1518-002556, 1518-002557, 1518-002558, 1518-002559, 1518-002560, 1518-002561, 1518-002562, 1518-002563, 1518-002564, 1518-002565, 1518-002566, 1518-002567, 1518-002568, 1518-002569, 1518-002570, 1518-002571, 1518-002572, 1518-002573, 1518-002574, 1518-002575, 1518-002576, 1518-002577, 1518-002578, 1518-002579, 1518-002580, 1518-002581, 1518-002582, 1518-002583, 1518-002584, 1518-002585, 1518-002586, 1518-002587, 1518-002588, 1518-002589, 1518-002590, 1518-002591, 1518-002595, 1518-002606, 1518-002607, 1518-002608, 1518-002609, 1518-002610, 1518-002611, 1518-002612, 1518-002613, 1518-002614,  and 1518-005744.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution: (USA) Nationwide distribution, and the countries of ARGENTINA, BAHRAIN, BOLIVIA, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, ECUADOR, EGYPT, FRANCE, INDIA, MEXICO, NICARAGUA, PARAGUAY, PERU, POLAND, Slovakia, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, CHINA, THAILAND, UNITED ARAB EMIRATES, U.K., URUGUAY, VENEZUELA, and VIETNAM.
  • Product Description
    Ultraview Spacelabs Medical Multigas Analyzer, Model 91518, V1.00.14. || Spacelabs Healthcare Llc. || Monitors gas concentrations and alerts clinical personnel when the concentration of anesthetic agents, oxygen, carbon dioxide, or nitrous oxide falls outside of defined limits. The anesthetic agent being administered is automatically identified.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Spacelabs Healthcare, Llc, 5150 220th Ave Se, Issaquah WA 98029-6834
  • Manufacturer Parent Company (2017)
  • Source
    USFDA